Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al–4 mm Cu HVL)
Authors:
- J E BurnsJBJ E Burns
- A G GreenerAGA G Greener
- R M HarrisonRHR M Harrison
- C MorettiCMC Moretti
- A E NahumANA E Nahum
- K E RosserKRK E Rosser
- 25 May 2005
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (12) , 2739-2748
Abstract
This addendum to the code of practice for the determination of absorbed dose for x-rays below 300 kV has recently been approved by the IPEM and introduces three main changes: (i) Due to a lack of available data the original code recommended a value of unity for k(ch) in the very-low-energy range (0.035-1.0 mm Al HVL). A single table of k(ch) values, ranging from 1.01 to 1.07, applicable to both designated chamber types is now presented. (ii) For medium-energy x-rays (0.5-4 mm Cu HVL) methods are given to determine the absorbed dose to water either at 2 cm depth or at the surface of a phantom depending on clinical needs. Determination of the dose at the phantom surface is derived from an in-air measurement and by extending the low-energy range up to 4 mm Cu HVL. Relevant backscatter factors and ratios of mass energy absorption coefficients are given in the addendum. (iii) Relative dosimetry: although not normally forming part of a dosimetry code of practice a brief review of the current literature on this topic has been added as an appendix. This encompasses advice on techniques for measuring depth doses, applicator factors for small field sizes, dose fall off with increasing SSD and choice of appropriate phantom materials and ionization chambers.Keywords
This publication has 15 references indexed in Scilit:
- Air kerma calibration factors and chamber correction values for PTW soft x-ray, NACP and Roos ionization chambers at very low x-ray energiesPhysics in Medicine & Biology, 2001
- AAPM protocol for 40–300 kV x‐ray beam dosimetry in radiotherapy and radiobiologyMedical Physics, 2001
- Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beamsPhysics in Medicine & Biology, 1997
- Low energy X-ray depth dose data for use in radiotherapy--comments on the review of BJR Supplement 17.British Journal of Radiology, 1997
- The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al - 4 mm Cu HVL; 10 - 300 kV generating potential)Physics in Medicine & Biology, 1996
- Dependence of the photon backscatter factor for water on source-to-phantom distance and irradiation field sizePhysics in Medicine & Biology, 1990
- Experimentally determined backscatter factors for X-rays generated at voltages between 16 and 140 kVPhysics in Medicine & Biology, 1989
- Contamination of ionization chambers by talcum powderMedical Physics, 1988
- A solid water phantom material for radiotherapy x-ray and γ-ray beam calibrationsMedical Physics, 1982
- CorrespondenceBritish Journal of Radiology, 1949