In vitro hemocompatibility testing: The importance of fresh blood
Open Access
- 12 February 2016
- journal article
- research article
- Published by American Vacuum Society in Biointerphases
- Vol. 11 (2) , 029802
Abstract
The use of unactivated blood for hemocompatibility testing is essential to obtain reliable results. Here, the authors study the influence of heparinized whole blood storage time and temperature on blood activation and evaluate the importance of initiating hemocompatibility tests within 4 h of blood collection. Blood from healthy volunteers was collected and analyzed with minimal delay, after 30 min and after 60 min of storage at room temperature, 30 or 37 °C. In addition, blood was analyzed after 1, 2, or 4 h of storage at room temperature. Platelet count, mean platelet volume, platelet binding capacity to collagen and thromboxane B2 were measured to assess platelet function, complement complex C5b-9 and elastase were measured to assess activation of the inflammatory response system, and thrombin-antithrombin III was measured to assess activation of the coagulation system. Furthermore, free hemoglobin was measured in platelet poor plasma as an indicator for red blood cell damage. The authors found that storage at 30 °C significantly increased platelet and coagulation activity after 60 min and storage at 37 °C significantly increased platelet, coagulation, and white blood cell activity after 60 min. Storage at room temperature significantly decreased platelet binding to collagen after 4 h and increased platelet activity after 1 h onward and white blood cell activity after 4 h. Their results show that short-term storage of heparinized whole blood significantly influences biomarkers over time, especially at 30 and 37 °C compared to room temperature. However, blood stored at room temperature for 4 h is also affected. In particular, platelet function and white blood cell activity are significantly influenced after 4 h of stationary storage at room temperature; therefore, the authors propose that hemocompatibility tests should be initiated well within 4 h of blood collection, preferably within 2 h.Keywords
This publication has 14 references indexed in Scilit:
- Effect of Blood Sampling, Processing, and Storage on the Measurement of Complement Activation BiomarkersAmerican Journal of Clinical Pathology, 2015
- Development and hemocompatibility testing of nitric oxide releasing polymers using a rabbit model of thrombogenicityJournal of Biomaterials Applications, 2014
- Changes in platelet morphology and function during 24 hours of storageClinical Hemorheology and Microcirculation, 2014
- Are there sufficient standards for the in vitro hemocompatibility testing of biomaterials?Biointerphases, 2013
- Obstacles in Haemocompatibility TestingScientifica, 2013
- In vitro haemocompatibility of a novel bioprosthetic total artificial heartEuropean Journal of Cardio-Thoracic Surgery, 2012
- Comparison of Modified Chandler, Roller Pump, and Ball Valve Circulation Models forIn VitroTesting in High Blood Flow Conditions: Application in Thrombogenicity Testing of Different Materials for Vascular ApplicationsInternational Journal of Biomaterials, 2012
- The Platelet Storage LesionClinics In Laboratory Medicine, 2010
- The platelet storage lesionTransfusion and Apheresis Science, 2009
- A Colorimetric Method for the Measurement of Platelet Adhesion in Microtiter PlatesAnalytical Biochemistry, 1994