Towards reference dosimetry for the MR-linac: magnetic field correction of the ionization chamber reading
Authors:
- K SmitKSK Smit
- B van AsselenBAB van Asselen
- J G M KokJKJ G M Kok
- A H L AalbersAAA H L Aalbers
- J J W LagendijkJLJ J W Lagendijk
- B W RaaymakersBRB W Raaymakers
- 12 August 2013
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 58 (17) , 5945-5957
Abstract
In the UMC Utrecht a prototype MR-linac has been installed. The system consists of a 6 MV Elekta (Crawley, UK) linear accelerator and a 1.5 T Philips (Best, The Netherlands) Achieva MRI system. This paper investigates the feasibility to correct the ionization chamber reading for the magnetic field within the dosimetry calibration method described by Almond et al (1999 Med. Phys. 26 1847–70). Firstly, the feasibility of using an ionization chamber in an MR-linac was assessed by investigating possible influences of the magnetic field on NE2571 Farmer-type ionization chamber characteristics: linearity, repeatability, orientation in the magnetic field; and AAPM TG51 correction factor for voltage polarity and ion recombination. We found that these AAPM correction factors for the NE2571 chamber were not influenced by the magnetic field. Secondly, the influence of the permanent 1.5 T magnetic field on the NE2571 chamber reading was quantified. The reading is influenced by the magnetic field; therefore, a correction factor has been added. For the standardized setup used in this paper, the NE2571 chamber reading increases by 4.9% (± 0.2%) due to the transverse 1.5 T magnetic field. Dosimetry measurements in an MR-linac are feasible, if a setup-specific magnetic field correction factor (P1.5 T) for the charge reading is introduced. For the setup investigated in this paper, the P1.5 T has a value of 0.953.Keywords
This publication has 12 references indexed in Scilit:
- Integrating a 1.5 T MRI scanner with a 6 MV accelerator: proof of conceptPhysics in Medicine & Biology, 2009
- Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamberPhysics in Medicine & Biology, 2009
- MRI/linac integrationRadiotherapy and Oncology, 2008
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: impact of the surface orientation on the entrance and exit dose due to the transverse magnetic fieldPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose deposition in a transverse magnetic fieldPhysics in Medicine & Biology, 2004
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration servicePhysics in Medicine & Biology, 1990
- The Saturation Curve for Ionization Measurements in Pulsed Radiation BeamsBritish Journal of Radiology, 1952
- On the Theory of RecombinationPhysical Review B, 1940