Analytical Screening of Low-cost CO₂ Sensors for Clinical Infection Control Applications
- 25 March 2026
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Clinical Engineering
- Vol. 51 (2) , 70-76
Abstract
We evaluated the accuracy and reliability of 12 low-cost CO₂ sensors that are commercially available and promoted for infection control in indoor settings, considering their potential use in clinical environments. A stepwise CO₂ injection protocol and alcohol exposure tests were conducted in a controlled acrylic chamber. Our results showed that only 3 out of 12 sensors (25%) responded reliably to actual CO₂ concentrations and could be calibrated with acceptable accuracy, whereas 67% failed to detect CO₂ but falsely responded to alcohol vapors, frequently present in health care environments. These findings highlight the need for staff development educators to promote accurate sensor selection and propose a practical 3-step screening method that clinical engineers can use without specialized equipment.Keywords
This publication has 16 references indexed in Scilit:
- CO2 concentration as an indicator of indoor ventilation performance to control airborne transmission of SARS-CoV-2Journal of Infection and Public Health, 2023
- Monitoring carbon dioxide to quantify the risk of indoor airborne transmission of COVID-19Flow, 2021
- Possible aerosol transmission of COVID-19 associated with an outbreak in an apartment in Seoul, South Korea, 2020International Journal of Infectious Diseases, 2021
- Mechanistic transmission modeling of COVID-19 on the Diamond Princess cruise ship demonstrates the importance of aerosol transmissionProceedings of the National Academy of Sciences, 2021
- Estimation of airborne viral emission: Quanta emission rate of SARS-CoV-2 for infection risk assessmentEnvironment International, 2020
- Intelligent Calibration and Virtual Sensing for Integrated Low-Cost Air Quality SensorsIEEE Sensors Journal, 2020
- Can commercial low-cost sensor platforms contribute to air quality monitoring and exposure estimates?Environment International, 2017
- Evaluating the performance of low cost chemical sensors for air pollution researchFaraday Discussions, 2015
- Risk of indoor airborne infection transmission estimated from carbon dioxide concentrationIndoor Air, 2003
- AIRBORNE SPREAD OF MEASLES IN A SUBURBAN ELEMENTARY SCHOOLAmerican Journal of Epidemiology, 1978