Abstract
Pencil beam scanning proton therapy allows the delivery of highly conformal dose distributions by delivering several thousand pencil beams. These beams have to be individually optimised and accurately delivered requiring a significant quality assurance workload. In this work we describe a toolkit for independent dose calculations developed at Paul Scherrer Institut which allows for dose reconstructions at several points in the treatment workflow. Quality assurance based on reconstructed dose distributions was shown to be favourable to pencil beam by pencil beam comparisons for the detection of delivery uncertainties and estimation of their effects. Furthermore the dose reconstructions were shown to have a sensitivity of the order of or higher than the measurements currently employed in the clinical verification procedures. The design of the independent dose calculation tool allows for a high modifiability of the dose calculation parameters (e.g. depth dose profiles, angular spatial distributions) allowing for a safe environment outside of the clinical treatment planning system for investigating the effect of such parameters on the resulting dose distributions and thus distinguishing between different contributions to measured dose deviations. The presented system could potentially reduce the amount of patient-specific quality assurance measurements which currently constitute a bottleneck in the clinical workflow.Keywords
This publication has 20 references indexed in Scilit:
- Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assuranceMedical Physics, 2013
- Patient-Specific Quality Assurance for Prostate Cancer Patients Receiving Spot Scanning Proton Therapy Using Single-Field Uniform DoseInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- More than 10 years experience of beam monitoring with the Gantry 1 spot scanning proton therapy facility at PSIMedical Physics, 2009
- Retrospective IMRT Dose Reconstruction Based on Cone-Beam CT and MLC Log-FileInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Replacing Pretreatment Verification With In Vivo EPID Dosimetry for Prostate IMRTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Monte Carlo dose calculations for spot scanned proton therapyPhysics in Medicine & Biology, 2005
- Treatment planning and verification of proton therapy using spot scanning: Initial experiencesMedical Physics, 2004
- Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facilityMedical Physics, 2004
- Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniquesMedical Physics, 2000
- The calibration of CT Hounsfield units for radiotherapy treatment planningPhysics in Medicine & Biology, 1996