A Computer-Aided Diagnosis for Locating Abnormalities in Bone Scintigraphy by a Fuzzy System With a Three-Step Minimization Approach
Authors:
- Tang-Kai YinTYTang-Kai Yin
- Nan-Tsing ChiuNCNan-Tsing Chiu
- 4 May 2004
- journal article
- clinical trial
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 23 (5) , 639-654
Abstract
Bone scintigraphy is an effective method to diagnose bone diseases such as bone tumors. In the scintigraphic images, bone abnormalities are widely scattered on the whole body. Conventionally, radiologists visually check the whole-body images and find the distributed abnormalities based on their expertise. This manual process is time-consuming and it is not unusual to miss some abnormalities. In this paper, a computer-aided diagnosis (CAD) system is proposed to assist radiologists in the diagnosis of bone scintigraphy. The system will provide warning marks and abnormal scores on some locations of the images to direct radiologists' attention toward these locations. A fuzzy system called characteristic-point-based fuzzy inference system (CPFIS) is employed to implement the diagnosis system and three minimizations are used to systematically train the CPFIS. Asymmetry and brightness are chosen as the two inputs to the CPFIS according to radiologists' knowledge. The resulting CAD system is of a small-sized rule base such that the resulting fuzzy rules can be not only easily understood by radiologists, but also matched to and compared with their expert knowledge. The prototype CAD system was tested on 82 abnormal images and 27 normal images. We employed free-response receiver operating characteristics method with the mean number of false positives (FPs) and the sensitivity as performance indexes to evaluate the proposed system. The sensitivity is 91.5% (227 of 248) and the mean number of FPs is 37.3 per image. The high sensitivity and moderate numbers of FP marks per image shows that the proposed method can provide an effective second-reader information to radiologists in the diagnosis of bone scintigraphy.Keywords
This publication has 30 references indexed in Scilit:
- A characteristic-point-based fuzzy inference system2002
- Designing fuzzy inference systems from data: An interpretability-oriented reviewIEEE Transactions on Fuzzy Systems, 2001
- Computerized radiographic mass detection. II. Decision support by featured database visualization and modular neural networksIEEE Transactions on Medical Imaging, 2001
- A computer-aided diagnosis system for digital mammograms based on fuzzy-neural and feature extraction techniquesIEEE Transactions on Information Technology in Biomedicine, 2001
- A CAD system for the automatic detection of clustered microcalcifications in digitized mammogram filmsIEEE Transactions on Medical Imaging, 2000
- A novel approach to microcalcification detection using fuzzy logic techniqueIEEE Transactions on Medical Imaging, 1998
- Computer-aided diagnosis: a neural-network-based approach to lung nodule detectionIEEE Transactions on Medical Imaging, 1998
- Consideration of the method of image diagnosis with respect to frontal lobe atrophyIEEE Transactions on Nuclear Science, 1996
- Soft computing and fuzzy logicIEEE Software, 1994
- Neural-network-based fuzzy logic control and decision systemIEEE Transactions on Computers, 1991