Bio-WiTel: A Low-Power Integrated Wireless Telemetry System for Healthcare Applications in 401–406 MHz Band of MedRadio Spectrum
- 14 December 2016
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Biomedical and Health Informatics
- Vol. 22 (2) , 483-494
Abstract
This paper presents a low-power integrated wireless telemetry system (Bio-WiTel) for healthcare applications in 401–406 MHz frequency band of medical device radiocommunication (MedRadio) spectrum. In this paper, necessary design considerations for telemetry system for short-range (upto 3 m) communication of biosignals are presented. These considerations help greatly in making important design decisions, which eventually lead to a simple, low power, robust, and reliable wireless system implementation. Transmitter (TX) and receiver (RX) of Bio-WiTel system have been fabricated in 180 nm mixed mode CMOS technology. While radiating −18 dBm output power to a 50 $\Omega$ antenna, the packaged TX IC consumes 250 μW power in 100% on state from 1 V supply, whereas the RX IC consumes 990 μW power from 1.8 V supply with a sensitivity of −75 dBm. Measurement results show that TX fulfils the spectral mask requirement at a maximum data rate of 72 kb/s. The measured bit error rate (BER) of RX is less than $10^{-4}$ for a data rate of 200 kb/s. The proposed Bio-WiTel system is tested successfully in home and hospital environments for the communication of electrocardiogram and photoplethysmogram signals at a data rate of 57.6 kb/s with a measured BER of $^{-4}$ for a maximum distance of 3 m.Keywords
Funding Information
- Ministry of Electronics and Information technology
- Government of India
This publication has 22 references indexed in Scilit:
- A 0.33 nJ/bit IEEE802.15.6/Proprietary MICS/ISM Wireless Transceiver With Scalable Data Rate for Medical Implantable ApplicationsIEEE Journal of Biomedical and Health Informatics, 2015
- Integrated Circuits and Electrode Interfaces for Noninvasive Physiological MonitoringIEEE Transactions on Biomedical Engineering, 2014
- A 0.7-V 17.4-/spl mu/W 3-Lead Wireless ECG SoCIEEE Transactions on Biomedical Circuits and Systems, 2013
- A 0.5 ${\rm cm}^{3}$ Four-Channel 1.1 mW Wireless Biosignal Interface With 20 m RangeIEEE Transactions on Biomedical Circuits and Systems, 2013
- Nurse–Patient Ratios as a Patient Safety StrategyAnnals of Internal Medicine, 2013
- A Batteryless 19 $\mu$W MICS/ISM-Band Energy Harvesting Body Sensor Node SoC for ExG ApplicationsIEEE Journal of Solid-State Circuits, 2013
- A 350 $\mu$W CMOS MSK Transmitter and 400 $\mu$W OOK Super-Regenerative Receiver for Medical Implant CommunicationsIEEE Journal of Solid-State Circuits, 2009
- The Correlation Coefficient: An OverviewCritical Reviews in Analytical Chemistry, 2006
- Jitter and phase noise in ring oscillatorsIEEE Journal of Solid-State Circuits, 1999
- A general theory of phase noise in electrical oscillatorsIEEE Journal of Solid-State Circuits, 1998