A differentially pumped electron-energy-loss spectrometer with multichannel detector for time-resolved studies at intermediate ambient pressures
Authors:
- 1 February 1992
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 63 (2) , 1652-1670
Abstract
The design, construction, and operation of a high‐resolution electron‐energy‐loss spectrometer featuring a position‐sensitive resistive anode detector and differential pumping are described. The position‐sensitive detector provides a 66‐fold increase in acquisition speed over a single‐channel spectrometer. The differential pumping provides pressure isolation of 105 between the spectrometer elements, except the lenses, and the sample region. An optically isolated multicomputer‐based control and data acquisition system provides complete spectrometer control and automated sample cleaning and characterization. The spectrometer, together with a differentially pumped supersonic molecular beamline, permits in situ measurements of the interaction of a wide range of gas molecules with solid surfaces.Keywords
This publication has 50 references indexed in Scilit:
- Contributions of(v=0) and(v=1) to the dissociative adsorption of hydrogen on Cu(110)Physical Review Letters, 1990
- Effect of vibrational energy on the dissociative chemisorption of N2 on Fe(111)The Journal of Chemical Physics, 1987
- Direct, real-time measurement of adsorption and desorption kinetics on Cu(100)Journal of Electron Spectroscopy and Related Phenomena, 1986
- High-resolution He-scattering apparatus for gas–surface interaction studiesReview of Scientific Instruments, 1986
- Resonant Sticking at SurfacesPhysical Review Letters, 1986
- Effect of translational energy on the molecular chemisorption of CO on Ni(111): Implications for the dynamics of the chemisorption processThe Journal of Chemical Physics, 1986
- Observation of Direct Inelastic Scattering in the Presence of Trapping-Desorption Scattering: Xe on Pt(111)Physical Review Letters, 1979
- High resolution electron energy loss spectrometer for vibrational surface studiesJournal of Vacuum Science and Technology, 1979
- The influence of magnetic and electric fringing fields on the trajectories of charged particlesNuclear Instruments and Methods, 1965
- Ablenkung von Kathoden- und Kanalstrahlen am Rande eines Kondensators, dessen Streufeld durch eine Blende begrenzt istThe European Physical Journal A, 1935