The Failure of Negative Pressure Rewarming (Thermostat™) to Accelerate Recovery from Mild Hypothermia in Postoperative Surgical Patients

Authors:
  • CSCharles E. Smith
    Charles E. Smith
  • APAli Parand
    Ali Parand
  • APAlfred C. Pinchak
    Alfred C. Pinchak
  • JHJoan F. Hagen
    Joan F. Hagen
  • DHDon E. Hancock
    Don E. Hancock

Abstract
The ThermostatTM device (Aquarius Medical Corp., Phoenix, AZ) is used in a new technique to accelerate recovery from hypothermia by mechanically distending blood vessels in the hand, thereby increasing transfer of exogenous heat to the body core. We evaluated the use of the ThermostatTM device in patients with mild postoperative hypothermia (TM, which consisted of a hypothermia warming mitt/seal and thermal exchange chamber for 60 min, and 2) conventional treatment, which consisted of warm blankets and/or radiant heat. Of the 191 patients enrolled, 60 (31%) developed hypothermia and were randomized to receive the ThermostatTM (n = 30) or conventional methods (n = 30). Fourteen patients in the ThermostatTM group and 17 patients in the conventional group rewarmed to 36°C before discharge from the recovery room (P is not significant). There were no differences in vital signs, rewarming time, time to discharge from the recovery room, or postoperative temperature between groups. We conclude that patients with mild postoperative hypothermia rewarmed in a similar fashion, regardless of whether the ThermostatTM or conventional methods were used. Implications We found that a commercially available negative pressure rewarming device (ThermostatTM; Aquarius Medical Corp., Phoenix, AZ) was not effective in accelerating rewarming in postoperative hypothermic surgical patients after general anesthesia.