Room temperature, high-resolution X-ray spectroscopy with silicon drift chambers
Authors:
- E. PinottiEPE. Pinotti
- A. LongoniALA. Longoni
- M. GambelliMGM. Gambelli
- L. StruderLSL. Struder
- P. LechnerPLP. Lechner
- C.V. ZanthierCZC.V. Zanthier
- H. KranerHKH. Kraner
- 1 February 1995
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 42 (1) , 12-16
Abstract
The development of detectors for high resolution room temperature X-ray spectroscopy represents a relevant progress in many fields of application, mainly in out-of-laboratory environments. A new type of silicon detector, the semiconductor drift chamber (SDC), allows one to obtain at room temperature, or with a moderate cooling, a resolution comparable to that obtained at liquid nitrogen temperature with traditional detectors of the same active area. The key feature of the SDC's is the very low output capacitance (about 100 fF) independent of the active area of the device. This feature, together with a good capacitive matching between the detector and the first stage of amplification, leads to high values of the resolution at short shaping times. We tested a simple 6-mm2 cylindrical SDC at room temperature and at -20°C (easily obtainable with electrical cooling), by using a specifically designed, low capacitance, JFET as the input transistor of the preamplifier. With a 55Fe source, we measured an equivalent noise charge (ENC) of 34 e- RMS and 27 e- RMS at room temperature and at -20°C respectively. To our knowledge these are presently the best values obtained for the same active area near room temperatureKeywords
This publication has 10 references indexed in Scilit:
- A method for the determination of the noise parameters in preamplifying systems for semiconductor radiation detectorsReview of Scientific Instruments, 1993
- The pn-CCD on-chip electronicsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993
- Suboptimal filtering of 1/ƒ-noise in detector charge measurementsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1990
- The MPI/AIT X-ray imager (MAXI) — High speed pn CCDs for X-ray detectionNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1990
- Instability of the behaviour of high resistivity silicon detectors due to the presence of oxide chargesNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1990
- Progress in semiconductor drift detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1986
- Processing the signals from solid-state detectors in elementary-particle physicsLa Rivista del Nuovo Cimento, 1986
- Semiconductor drift chambers for position and energy measurementsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1985
- Semiconductor drift chamber — An application of a novel charge transport schemeNuclear Instruments and Methods in Physics Research, 1984
- Field Effect Transistors for Charge AmplifiersIEEE Transactions on Nuclear Science, 1973