In-beam PET imaging for on-line adaptive proton therapy: an initial phantom study
Authors:
- Xishan SunXSXishan Sun
- Kai LouKLKai Lou
- Xiaorong R ZhuXZXiaorong R Zhu
- Dragon MirkovicDMDragon Mirkovic
- Falk PoenischFPFalk Poenisch
- David GrosshansDGDavid Grosshans
- 30 May 2014
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (13) , 3373-3388
Abstract
We developed and investigated a positron emission tomography (PET) system for use with on-line (both in-beam and intra-fraction) image-guided adaptive proton therapy applications. The PET has dual rotating depth-of-interaction measurable detector panels by using solid-state photomultiplier (SSPM) arrays and LYSO scintillators. It has a 44 mm diameter trans-axial and 30 mm axial field-of-view (FOV). A 38 mm diameter polymethyl methacrylate phantom was placed inside the FOV. Both PET and phantom axes were aligned with a collimated 179.2 MeV beam. Each beam delivered ∼50 spills (0.5 s spill and 1.5 s inter-spill time, 3.8 Gy at Bragg peak). Data from each beam were acquired with detectors at a given angle. Nine datasets for nine beams with detectors at nine different angles over 180° were acquired for full-tomographic imaging. Each dataset included data both during and 5 min after irradiations. The positron activity-range was measured from the PET image reconstructed from all nine datasets and compared to the results from simulated images. A (22)Na disc-source was also imaged after each beam to monitor the PET system's performance. PET performed well except for slight shifts of energy photo-peak positions (<1%) after each beam, due mainly to the neutron exposure of SSPM that increased the dark-count noise. This minor effect was corrected offline with a shifting 350-650 keV energy window for each dataset. The results show a fast converging of activity-ranges measured by the prototype PET with high sensitivity and uniform resolution. Sub-mm activity-ranges were achieved with minimal 6 s acquisition time and three spill irradiations. These results indicate the feasibility of PET for intra-fraction beam-range verification. Further studies are needed to develop and apply a novel clinical PET system for on-line image-guided adaptive proton therapy.Keywords
This publication has 39 references indexed in Scilit:
- Development of an Eight-Channel Time-Based Readout ASIC for PET ApplicationsIEEE Transactions on Nuclear Science, 2011
- A Monte Carlo pencil beam scanning model for proton treatment plan simulation using GATE/GEANT4Physics in Medicine & Biology, 2011
- The reliability of proton-nuclear interaction cross-section data to predict proton-induced PET images in proton therapyPhysics in Medicine & Biology, 2011
- In vivo verification of proton beam path by using post‐treatment PET/CT imagingMedical Physics, 2009
- Absolute dose reconstruction in proton therapy using PET imaging modality: feasibility studyPhysics in Medicine & Biology, 2009
- The Feasibility of In-Beam PET for Therapeutic Beams of$^3$HeIEEE Transactions on Nuclear Science, 2006
- On the detector arrangement for in-beam PET for hadron therapy monitoringPhysics in Medicine & Biology, 2006
- Positron Emission Tomography2005
- Charged hadron tumour therapy monitoring by means of PETNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- List-mode maximum-likelihood reconstruction applied to positron emission mammography (PEM) with irregular samplingIEEE Transactions on Medical Imaging, 2000