Comparison of methods for quantitative biomolecular interaction analysis
Authors:
Open Access
- 10 September 2021
- journal article
- research article
- Published by Springer Nature in Analytical and Bioanalytical Chemistry
- Vol. 414 (1) , 661-673
Abstract
In order to perform good kinetic experiments, not only the experimental conditions have to be optimized, but the evaluation procedure as well. The focus of this work is the in-depth comparison of different approaches and algorithms to determine kinetic rate constants for biomolecular interaction analysis (BIA). The different algorithms are applied not only to flawless simulated data, but also to real-world measurements. We compare five mathematical approaches for the evaluation of binding curves following pseudo-first-order kinetics with different noise levels. In addition, reflectometric interference spectroscopy (RIfS) measurements of two antibodies are evaluated to determine their binding kinetics. The advantages and disadvantages of the individual approach will be investigated and discussed in detail. In summary, we will raise awareness on how to evaluate and judge results from BIA by using different approaches rather than having to rely on “black box” closed (commercial) software packages.Keywords
Funding Information
- Eberhard Karls Universität Tübingen
This publication has 35 references indexed in Scilit:
- Grading the commercial optical biosensor literature—Class of 2008: ‘The Mighty Binders’Journal of Molecular Recognition, 2009
- Survey of the year 2005 commercial optical biosensor literatureJournal of Molecular Recognition, 2006
- Monitoring an antibody affinity chromatography with a label-free optical biosensor techniqueJournal of Immunological Methods, 2004
- Probing the Binding of Coumarins and Cyclothialidines to DNA GyraseBiochemistry, 1999
- An integrated system for optical biomolecular interaction analysisBiosensors and Bioelectronics, 1997
- Experimental design for kinetic analysis of protein-protein interactions with surface plasmon resonance biosensorsJournal of Immunological Methods, 1997
- Kinetic analysis of biosensor data: elementary tests for self-consistencyTrends in Biochemical Sciences, 1996
- Direct observation of affinity reactions by reflected-mode operation of integrated optical grating couplerSensors and Actuators B: Chemical, 1996
- Biosensing with surface plasmon resonance — how it all startedBiosensors and Bioelectronics, 1995