Mechanical Evaluation of a 4-mm Cancellous “Rescue” Screw in Osteoporotic Cortical Bone: A Cadaveric Study
Authors:
- Sandeep P SoinSSSandeep P Soin
- Trevor A KnightTKTrevor A Knight
- Simon C MearsSMSimon C Mears
- Stephen M BelkoffSBStephen M Belkoff
- 1 June 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 24 (6) , 379-382
Abstract
Obtaining sufficient fracture fixation in osteoporotic bone is challenging. The purposes of the current study were (1) to compare the pullout strength of a 4-mm cancellous screw (cancellous screw group) with that of a 3.5-mm cortical screw (cortical screw group), and (2) to measure the pullout strength of a 4-mm cancellous screw placed as a rescue screw (rescue screw group) in a stripped 3.5-mm cortical screw (stripped screw group) hole while controlling for bone density and cortical thickness. We inserted 4 screws, one from each experimental group, into 11 osteoporotic cadaveric radii, while recording the insertion torque. Radii were mounted on a servohydraulic testing machine, and each screw was pulled out at a rate of 5 mm/min. Pullout strength was recorded. The effects of cortical thickness (near, far, and total), bone density, insertion torque, and the experimental screw group (cortical, cancellous, stripped, and rescue screw groups) on pullout strength were analyzed using multiple linear regression with random effects. Statistical significance was set at P < 0.05. There was no significant difference in pullout strength between the cortical and cancellous screw groups. The rescue screw group had significantly less pullout strength than did the cortical and cancellous screws, and only partly increased pullout strength compared with stripped screws. Bone density significantly affected pullout strength, but insertion torque and cortical thickness were not significant covariates. There seems to be no advantage in using a cancellous screw over a cortical screw in bicortical fixation in osteoporotic bone. Although the rescue screw provided greater pullout strength than the stripped screw, it is unknown if the purchase it provides is clinically sufficient.Keywords
This publication has 15 references indexed in Scilit:
- Locked Plating in Orthopaedic Trauma: A Clinical UpdateJournal of the American Academy of Orthopaedic Surgeons, 2008
- Factors affecting the stability of screws in human cortical osteoporotic boneThe Journal of Bone and Joint Surgery. British volume, 2007
- Combination of Hook Plate and Tibial Pro-Fibular Screw Fixation of Osteoporotic Fractures: A Clinical Evaluation of Operative StrategyFoot & Ankle International, 2005
- Mechanical behavior of screws in normal and osteoporotic boneOsteoporosis International, 2004
- An Augmented Fixation Method for Distal Fibular Fractures in Elderly Patients: A Biomechanical EvaluationFoot & Ankle International, 2004
- Use of a synthetic bone void filler to augment screws in osteopenic ankle fracture fixationArchives of orthopaedic and trauma surgery, 2004
- Mechanical analysis of the bone to plate interface of the LC-DCP and of the PC-FIX on human femoraInjury, 2000
- Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws and the friction between plate and boneInjury, 2000
- The Reinforcement of Cancellous Bone Screws with Calcium Phosphate CementJournal of Orthopaedic Trauma, 1996
- Holding power of the 4.5 mm AO/ASIF cortex screw in cortical bone in relation to bone mineralInjury, 1993