Low Radioactive Material Screening and Background Control for the PandaX-4T Experiment

Authors:
  • ZQZhicheng Qian
    Zhicheng Qian
  • LSLin Si
    Lin Si
  • AAAbdusalam Abdukerim
    Abdusalam Abdukerim
  • ZBZihao Bo
    Zihao Bo
  • WCWei Chen
    Wei Chen
  • XCXun Chen
    Xun Chen
  • YCYunhua Chen
    Yunhua Chen
  • CCChen Cheng
    Chen Cheng
  • YCYunshan Cheng
    Yunshan Cheng
  • XCXiangyi Cui
    Xiangyi Cui
  • YFYingjie Fan
    Yingjie Fan
  • DFDeqing Fang
    Deqing Fang

Abstract
PandaX-4T is a ton-scale dark matter direct detection experiment using a dual-phase TPC technique at the China Jinping Underground Laboratory. Various ultra-low background technologies have been developed and applied to material screening for PandaX-4T, including HPGe gamma spectroscopy, ICP-MS, NAA, radon emanation measurement system, krypton assay station, and alpha detection system. Low background materials were selected to assemble the detector and surface treatment procedures were investigated to further suppress radioactive background. Combining measured results and Monte Carlo simulation, the total material background rates of PandaX-4T in the energy region of 1-25 keV$\rm{}_{ee}$ are estimated to be (9.6 $\pm$ 1.9) $\times \ 10^{-3}$ mDRU for electron recoil and (2.8 $\pm$ 0.6) $\times \ 10^{-4}$ mDRU for nuclear recoil. In addition, $^{222}$Rn and $^{nat}$Kr in the detector are estimated to be $\sim 6\ \mu$Bq/kg and <8 ppt.