On-chip integrated CMOS optical detection microsystem for spectrophotometric analyses in biological microfluidic systems
- 1 January 2005
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 3 (21635137) , 1133
Abstract
An integrated optical detection microsystem, which includes photodetectors and a light-to-frequency converter for readout, is designed and fabricated in a standard CMOS process without extra masks. This detection microsystem is designed for use in biological microsystems for fluids analysis. The application is in the low-cost concentration measurement of biomolecules in biological fluids, by the optical absorption in a part of the visible spectrum defined by the specific molecule. Signals proportional to the intensity of the light transmitted through the biological fluid are available at the output in the form of bit streams, which allows simple computer interfacing. The quantitative measurement of uric acid in urine is successfully demonstrated. The photodiode responsivity is 224 mA/W at /spl lambda/= 495 nm (the wavelength at which the uric acid has its absorption maximum). The optical system sensitivity is 1 kHz/Wm/sup -2/ at /spl lambda/ =670 nm (using the TLS230 from Texas Instruments as reference).Keywords
This publication has 4 references indexed in Scilit:
- Biological microsystem for measuring uric acid in biological fluidsSensors and Actuators A: Physical, 2003
- Optical properties of thin-film silicon-compatible materialsApplied Optics, 1997
- Clinical diagnostics opportunities for biosensors and bioelectronicsBiosensors and Bioelectronics, 1995
- Optical Constants of Silicon in the Region 1 to 10 evPhysical Review B, 1960