Verstrecktes Polyethylen (UHMWPE) - eine neue Werkstoffmodifikation zur Abriebsminimierung beim künstlichen Kniegelenksersatz - Stretched Polyethylene (UHMWPE) - a New Material Modification that Reduces Wear in Total Knee Arthroplasty
Authors:
- D. Ch. WirtzDWD. Ch. Wirtz
- E. SchopphoffESE. Schopphoff
- D. WeichertDWD. Weichert
- F. U. NiethardFNF. U. Niethard
- 1 January 2001
- journal article
- abstracts
- Published by Walter de Gruyter GmbH in Biomedical Engineering / Biomedizinische Technik
- Vol. 46 (12) , 338-342
Abstract
Polyethylene wear is a major cause of aseptic loosening of knee endoprostheses. With the aim of minimizing this mechanical wear, we stretch-modified the structure of polyethylene. From plates of UHMWPE, stretched PE samples were produced under defined conditions, and subsequently submitted to friction tests in a fluid environment (cycles 5 x 10(5), frequency 1.5 Hz, load 500 N, contact stress 10 MPa). After load testing, the stretched samples revealed a 70% reduced wear rate in comparison with non-stretched samples. Microscopic examination (light microscopy, scanning electron microscopy) showed abrasive scratches in all test samples, but only in the non-stretched specimens were signs of pitting found. Translating these results to unidirectional tractive rolling on the tibial plateau in the knee joint suggests that a significant reduction in polyethylene wear can be expected. However, further experimental investigations need to be carried out to confirm this highly promising possibility.Keywords