Relationship between aneurysm occlusion and flow diverting device oversizing in a rabbit model
- 11 November 2014
- journal article
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 8 (1) , 94-98
Abstract
Background and purpose Implanted, actual flow diverter pore density is thought to be strongly influenced by proper matching between the device size and parent artery diameter. The objective of this study was to characterize the correlation between device sizing, metal coverage, and the resultant occlusion of aneurysms following flow diverter treatment in a rabbit model. Methods Rabbit saccular aneurysms were treated with flow diverters (iso-sized to proximal parent artery, 0.5 mm oversized, or 1.0 mm oversized, respectively, n=6 for each group). Eight weeks after implantation, the angiographic degree of aneurysm occlusion was graded (complete, near-complete, or incomplete). The ostium of the explanted aneurysm covered with the flow diverter struts was photographed. Based on gross anatomic findings, the metal coverage and pore density at the ostium of the aneurysm were calculated and correlated with the degree of aneurysm occlusion. Results Angiographic results showed there were no statistically significant differences in aneurysm geometry and occlusion among groups. The mean parent artery diameter to flow diverter diameter ratio was higher in the 1.0 mm oversized group than in the other groups. Neither the percentage metal coverage nor the pore density showed statistically significant differences among groups. Aneurysm occlusion was inversely correlated with the ostium diameter, irrespective of the size of the device implanted. Conclusions Device sizing alone does not predict resultant pore density or metal coverage following flow diverter implantation in the rabbit aneurysm model. Aneurysm occlusion was not impacted by either metal coverage or pore density, but was inversely correlated with the diameter of the ostium.Keywords
This publication has 15 references indexed in Scilit:
- Analysis of flow changes in side branches jailed by flow diverters in rabbit modelsInternational Journal for Numerical Methods in Biomedical Engineering, 2014
- Strategy for analysis of flow diverting devices based on multi‐modality image‐based modelingInternational Journal for Numerical Methods in Biomedical Engineering, 2014
- Cellular Mechanisms of Aneurysm Occlusion after Treatment with a Flow DiverterRadiology, 2014
- Flow Diversion for Intracranial AneurysmsStroke, 2011
- Flow diverters for treatment of intracranial aneurysms: Current status and ongoing clinical trialsJournal of Clinical Neuroscience, 2011
- An Original Flow Diversion Device for the Treatment of Intracranial AneurysmsStroke, 2009
- Treatment of Rabbit Elastase-Induced Aneurysm Models by Flow Diverters: Development of Quantifiable Indexes of Device Performance Using Digital Subtraction AngiographyIEEE Transactions on Medical Imaging, 2009
- A New Endoluminal, Flow-Disrupting Device for Treatment of Saccular AneurysmsStroke, 2007
- Creation of Saccular Aneurysms in the RabbitAmerican Journal of Roentgenology, 2000
- Alteration of hemodynamics in aneurysm models by stenting: Influence of stent porosityAnnals of Biomedical Engineering, 1997