11.6 A multi-channel neural-recording amplifier system with 90dB CMRR employing CMOS-inverter-based OTAs with CMFB through supply rails in 65nm CMOS
Authors:
- 1 February 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 01936530,p. 1-3
Abstract
In addition to minimizing input-referred noise and lowering power consumption, a good multi-channel neural amplifier system should be able to significantly reject common-mode electrical interference (CMI). The dominant source of CMI comes from capacitive coupling of electrical mains supply line or EMGs onto neural tissues and can be as high as 100mVpp. Thus any neural recording setup needs a total common-mode rejection ratio (TCMRR) of at least 70dB for a minimum detectable neural signal of 5uVrms. However, multi-channel neural amplifiers are commonly implemented with a shared reference input whose input impedance is several times lower than that of corresponding signal inputs. This results in a large mismatch at the bipolar electrode-amplifier input interface. As analysed in Fig. 11.6.1, the TCMRR is significantly degraded below 70dB, independent of an amplifier's intrinsic CMRR (ICMRR). In this work, we report a micro-power, low-noise 16-channel neural amplifier that eliminates this impedance mismatch problem by using single-ended CMOS inverter-based LNAs for both the reference and signal inputs. Compared to conventional replica channel works, when operating at 1V supply, the LNAs can accommodate a large input CMI of up to 220mVpp through the use of a common-mode feedback (CMFB) loop implemented through the supply rails of the CMOS-inverter-based LNAs, which coincidentally leads to a high amplifier ICMRR.Keywords
This publication has 5 references indexed in Scilit:
- 64-Channel UWB Wireless Neural Vector Analyzer SOC With a Closed-Loop Phase Synchrony-Triggered NeurostimulatorIEEE Journal of Solid-State Circuits, 2013
- An implantable 455-active-electrode 52-channel CMOS neural probe2013
- Energy Efficient Low-Noise Neural Recording Amplifier With Enhanced Noise Efficiency FactorIEEE Transactions on Biomedical Circuits and Systems, 2011
- A 160μW 8-channel active electrode system for EEG monitoring2011
- Integrated circuit amplifiers for multi-electrode intracortical recordingJournal of Neural Engineering, 2009