B1 propagation properties of combined circularly polarised mode in ultra‐high‐field MRI
Authors:
Open Access
- 1 April 2016
- journal article
- research article
- Published by Institution of Engineering and Technology (IET) in Electronics Letters
- Vol. 52 (8) , 591-592
Abstract
In ultra‐high‐field magnetic resonance imaging (MRI), inhomogeneity of the radiofrequency transmit field (| B 1 + |) occurs due to wavelength shortening that results from interactions between the coil and the high dielectric subject. Even if a multiple transmit/receive (Tx/Rx) coil array is employed using the first‐order clockwise circularly polarised (|CP 1 + |) mode, | B 1 + | inhomogeneity is an insurmountable obstacle. A scheme utilising the CP 1 + mode for the odd‐numbered coil elements and the third‐order clockwise CP mode for the even‐numbered coil elements in the multiple‐element Tx/Rx coil array was investigated with the objective of increasing the | B 1 + | homogeneity at 300 MHz. The results of an electromagnetic simulation and a 7 T MRI experiment are presented and to illustrate the improvements achieved using the optimum CP mode combination is discussed.Keywords
This publication has 4 references indexed in Scilit:
- B 1 field comparison for RF coils in ultra‐high‐field MRIElectronics Letters, 2013
- Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channelsMagnetic Resonance in Medicine, 2008
- Transmit and receive transmission line arrays for 7 Tesla parallel imagingMagnetic Resonance in Medicine, 2005
- Reduction of wave phenomena in high-field MRI experiments using absorbing layersJournal of Magnetic Resonance, 2004