A correction algorithm to simultaneously control dual deformable mirrors in a woofer-tweeter adaptive optics system
Open Access
- 23 July 2010
- journal article
- research article
- Published by Optica Publishing Group in Optics Express
- Vol. 18 (16) , 16671-16684
Abstract
We present a direct slope-based correction algorithm to simultaneously control two deformable mirrors (DMs) in a woofer-tweeter adaptive optics system. A global response matrix was derived from the response matrices of each deformable mirror and the voltages for both deformable mirrors were calculated simultaneously. This control algorithm was tested and compared with a 2-step sequential control method in five normal human eyes using an adaptive optics scanning laser ophthalmoscope. The mean residual total root-mean-square (RMS) wavefront errors across subjects after adaptive optics (AO) correction were 0.128 ± 0.025 μm and 0.107 ± 0.033 μm for simultaneous and 2-step control, respectively (7.75-mm pupil). The mean intensity of reflectance images acquired after AO convergence was slightly higher for 2-step control. Radially-averaged power spectra calculated from registered reflectance images were nearly identical for all subjects using simultaneous or 2-step control. The correction performance of our new simultaneous dual DM control algorithm is comparable to 2-step control, but is more efficient. This method can be applied to any woofer-tweeter AO system.Keywords
This publication has 26 references indexed in Scilit:
- Imaging outer segment renewal in living human cone photoreceptorsOptics Express, 2010
- Accommodative lag and fluctuations when optical aberrations are manipulatedJournal of Vision, 2009
- Volumetric retinal imaging with ultrahigh-resolution spectral-domain optical coherence tomography and adaptive optics using two broadband light sourcesOptics Express, 2009
- Individual Variations in Human Cone Photoreceptor Packing Density: Variations with Refractive ErrorInvestigative Ophthalmology & Visual Science, 2008
- Visual performance in emmetropia and low myopia after correction of high-order aberrationsJournal of Vision, 2007
- In vivo fluorescence imaging of primate retinal ganglion cells and retinal pigment epithelial cellsOptics Express, 2006
- Adaptive optics scanning laser ophthalmoscopyOptics Express, 2002
- Improvement in retinal image quality with dynamic correction of the eye’s aberrationsOptics Express, 2001
- Zernike modal compensation analysis for an adaptive optics system using direct-gradient wavefront reconstruction algorithm1999
- Magnification and Visual Acuity in Refractive SurgeryArchives of Ophthalmology (1950), 1993