MR liver imaging and cholangiography in the presence of surgical metallic clips at 1.5 and 3 Tesla
Authors:
- 16 May 2006
- journal article
- hepatobiliary pancreas
- Published by Springer Nature in European Radiology
- Vol. 16 (10) , 2309-2316
Abstract
To evaluate whether clips from prior cholecystectomy impair image quality during magnetic resonance cholangiography (MRC) at 3 Tesla (T) compared with 1.5 T, surgical clips were embedded in a gel phantom and positioned at predefined distances from a fluid-filled tube designed to simulate the bile duct. The maximum clip distance was noted where susceptibility artifacts obscured the fluid-filled tube at 1.5 T and 3 T. Susceptibility artifact size was calculated for each sequence within each magnet class. In vivo analysis included 42 patients postcholecystectomy who underwent MRC at either 1.5 T or 3 T. In vitro, mean area of susceptibility artifacts was 104 mm2 on 3-T and 75 mm2 on 1.5-T MR imaging (MRI). While surgical clips within a 2-mm range impaired visualization of the fluid-filled tube on 1.5-T MRI, this range increased to 4 mm on 3-T MRI. In vivo, MRC image quality was impaired by susceptibility artifacts in three of 21 cases at 3 T and in two of 21 cases at 1.5 T. Overall, biliary pseudo-obstructions due to susceptibility artifacts from cholecystectomy surgical clips were not substantially more common on 3-T MRC in clinical practice, and patients with a history of prior cholecystectomy should not be excluded from a 3-T MRC.Keywords
This publication has 17 references indexed in Scilit:
- 3.0- Versus 1.5-T MR Cholangiography: A Pilot StudyAmerican Journal of Roentgenology, 2006
- Radio frequency versus susceptibility effects of small conductive implants — a systematic MRI study on aneurysm clips at 1.5 and 3 TMagnetic Resonance Imaging, 2005
- RF artifacts caused by metallic implants or instruments which get more prominent at 3 T: an in vitro studyMagnetic Resonance Imaging, 2005
- Whole-body MRI at high field: technical limits and clinical potentialEuropean Radiology, 2005
- 3.0 Tesla Magnetic Resonance Angiography of Endovascular Aortic Stent GraftsInvestigative Radiology, 2004
- Sensitivity of T2-weighted FSE sequences towards physiological iron depositions in normal brains at 1.5 and 3.0 TEuropean Radiology, 2004
- Magnetic Resonance CholangiopancreatographyNew England Journal of Medicine, 1999
- Visualization of magnetic resonance-compatible needles at 1.5 and 0.2 TeslaCardioVascular and Interventional Radiology, 1996
- Needle localization in MR-guided biopsy and aspiration: effects of field strength, sequence design, and magnetic field orientation.American Journal of Roentgenology, 1996
- Optimizing imaging parameters for MR evaluation of the spine with titanium pedicle screws.American Journal of Roentgenology, 1996