Anterior cervical locking plate-related complications; prevention and treatment recommendations
Authors:
- 12 May 2007
- journal article
- research article
- Published by Springer Nature in International Orthopaedics
- Vol. 32 (5) , 649-655
Abstract
A retrospective study evaluating complications in 2,233 consecutive patients of subaxial cervical disorders treated with an anterior cervical locking plate was performed, and recommendations for prevention and treatment were made. The average length of follow-up was 1.3 years. Any loosening or breaking of the plates and screws or malpositions that threatened tracheoesophageal or neurovascular structures were defined as the complications. There were 239 cases (10.7%) with different kinds of complications. The complications included oblique plating in 56 cases in which the screw could irritate the nerve root. Screws were driven into the disc space in four cases, which ultimately led to plate loosening. Screws penetrated the endplate or passed excessively close to it producing a triangle fracture in 19 cases. Loosening or breaking of the plate and the screw was found in 115 cases. These phenomena were always associated with non-union. Three oesophageal perforations occurred and conservative treatments proved effective. Finally, overlong plates impinged on the adjacent level in 14 cases and promoted disc degeneration ultimately leading to revision surgery. Good training and careful operation may help to decrease the complication rate. Most hardware complications are not symptomatic and can be treated conservatively. Only a few of them need immediate reoperation.Keywords
This publication has 22 references indexed in Scilit:
- Fracture of anterior cervical plate implant – report of two casesActa Neurochirurgica, 2005
- Management of the Cervical Esophagus and Hypofarinx Perforations Complicating Anterior Cervical Spine SurgerySpine, 2003
- The Significance of Hardware Failure in Anterior Cervical Plate FixationSpine, 1998
- Oesophageal perforation subsequent to anterior cervical spine screw/plate fixationSpinal Cord, 1995
- Anterior cervical fixation with the titanium locking screw-plate: A preliminary reportSurgical Neurology, 1994
- The Synthes Cervical Spine Locking Plate and Screw System in Anterior Cervical FusionTechniques in Orthopaedics, 1994
- Esophageal Penetration by an Anterior Cervical Fixation DeviceSpine, 1993
- Series of Ninety-Two Traumatic Cervical Spine Injuries Stabilized with Anterior ASIF Plate Fusion TechniqueSpine, 1991
- Cervical Stabilization by Plate and Bone FusionSpine, 1988
- The Anterior Cervical PlateSpine, 1983