A penalized-likelihood image reconstruction method for emission tomography, compared to postsmoothed maximum-likelihood with matched spatial resolution
Authors:
- 4 September 2003
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 22 (9) , 1042-1052
Abstract
Regularization is desirable for image reconstruction in emission tomography. A powerful regularization method is the penalized-likelihood (PL) reconstruction algorithm (or equivalently, maximum a posteriori reconstruction), where the sum of the likelihood and a noise suppressing penalty term (or Bayesian prior) is optimized. Usually, this approach yields position-dependent resolution and bias. However, for some applications in emission tomography, a shift-invariant point spread function would be advantageous. Recently, a new method has been proposed, in which the penalty term is tuned in every pixel to impose a uniform local impulse response. In this paper, an alternative way to tune the penalty term is presented. We performed positron emission tomography and single photon emission computed tomography simulations to compare the performance of the new method to that of the postsmoothed maximum-likelihood (ML) approach, using the impulse response of the former method as the postsmoothing filter for the latter. For this experiment, the noise properties of the PL algorithm were not superior to those of postsmoothed ML reconstruction.Keywords
This publication has 20 references indexed in Scilit:
- Object dependency of resolution and convergence rate in OSEM with filtering2005
- On estimating the variance of smoothed MLEM imagesIEEE Transactions on Nuclear Science, 2002
- Comparison of post-filtering and filtering between iterations for SPECT reconstructionIEEE Transactions on Nuclear Science, 1999
- High-resolution 3D Bayesian image reconstruction using the microPET small-animal scannerPhysics in Medicine & Biology, 1998
- Bayesian image reconstruction for emission tomography based on median root priorEuropean Journal of Nuclear Medicine and Molecular Imaging, 1997
- Objective assessment of image quality II Fisher information, Fourier crosstalk, and figures of merit for task performanceJournal of the Optical Society of America A, 1995
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- Recursive regularization filters: design, properties, and applicationsIEEE Transactions on Pattern Analysis and Machine Intelligence, 1991
- Maximum likelihood SPECT in clinical computation times using mesh-connected parallel computersIEEE Transactions on Medical Imaging, 1991
- Maximum Likelihood Reconstruction for Emission TomographyIEEE Transactions on Medical Imaging, 1982