Understanding the Radial Force of Stroke Thrombectomy Devices to Minimize Vessel Wall Injury: Mechanical Bench Testing of the Radial Force Generated by a Novel Braided Thrombectomy Assist Device Compared to Laser-Cut Stent Retrievers in Simulated MCA Vessel Diameters
Authors:
- 5 August 2019
- journal article
- research article
- Published by S. Karger AG in Interventional Neurology
- Vol. 8 (2-6) , 206-214
Abstract
Background: Recent reports have raised various concerns about the risk of vessel wall injury while withdrawing current laser-cut stent retrievers during active strut apposition to the vessel walls. The development of braided thrombectomy assist devices in conjunction with aspiration systems may be gentler on the fragile brain vessels and more optimized with regard to the radial force (RF) for vessel diameters of proximal (M1) and distal (M2) large vessel occlusions (LVOs). Methods: Mechanical bench testing of the RF was performed using a radial compression station mounted on a tensile testing machine. The total RF in newtons (N) generated in vessels with diameters ranging from 2.25 to 3 mm as seen in proximal LVOs (∼M1), and in vessel diameters ranging from 1.5 to 2.24 mm as seen in distal LVOs (∼M2), was measured. The outer diameter of each stent was recorded, and an RF ≤1 N was grouped as “low,” while an RF >1 N was grouped as “high” for this analysis. Results: The total RFs of all laser-cut stent retrievers were all higher in the simulated M2 vessels (>1 N) than in the M1 vessels (Conclusions: Novel braided thrombectomy assist devices in conjunction with aspiration systems have lower RFs than existing laser-cut stent retrievers in M1 and M2 vessel diameters. Further in vivo studies are needed to delineate the impact of lowering the RF on vessel wall integrity.Keywords
This publication has 28 references indexed in Scilit:
- Vessel Wall Magnetic Resonance Imaging in Acute Ischemic StrokeStroke, 2014
- Evolution of thrombectomy approaches and devices for acute stroke: a technical reviewJournal of NeuroInterventional Surgery, 2014
- Stent Retrievers in Acute Ischemic Stroke: Complications and Failures during the Perioperative PeriodAmerican Journal of Neuroradiology, 2013
- A Trial of Imaging Selection and Endovascular Treatment for Ischemic StrokeNew England Journal of Medicine, 2013
- Solitaire flow restoration device versus the Merci Retriever in patients with acute ischaemic stroke (SWIFT): a randomised, parallel-group, non-inferiority trialThe Lancet, 2012
- Trevo versus Merci retrievers for thrombectomy revascularisation of large vessel occlusions in acute ischaemic stroke (TREVO 2): a randomised trialThe Lancet, 2012
- Radial Force Measurements in Carotid Stents: Influence of Stent Design and Length of the LesionJournal of Vascular and Interventional Radiology, 2011
- A Comparison of Functional and Physical Properties of Self-Expanding Intracranial Stents [Neuroform3, Wingspan, Solitaire, Leo(+), Enterprise]min - Minimally Invasive Neurosurgery, 2011
- Intra-arterial versus intra-venous thrombolysis within and after the first 3 hours of stroke onsetArchives of Medical Science, 2010
- Guidelines for the Early Management of Adults With Ischemic StrokeCirculation, 2007