Design of an instrument for far-infrared microspectroscopy using a synchrotron radiation source
Authors:
- 1 January 1992
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 63 (1) , 1551-1554
Abstract
A design of a microspectrophotometric system using a synchrotron radiation (SR) source is described. The system covers a wide spectral range of 50–13000 cm−1, being under construction at the UVSOR BL6B beamline in the Institute for Molecular Science. Preliminary experiments in the mid‐infrared region (500–5000 cm−1) have qualitatively confirmed the theoretical calculation that the synchrotron radiation is more intense than a blackbody (T=1200 K) when a microspectrophotomeric technique is applied, which is due to natural collimation and high brilliance of SR source. The SR as an infrared source exhibits its advantage on measuring the spectra of small single crystals especially in the far‐infrared region.Keywords
This publication has 8 references indexed in Scilit:
- Infrared synchrotron radiation measurements at BrookhavenReview of Scientific Instruments, 1989
- Utilization of synchrotron radiation in the far-infrared region (invited)Review of Scientific Instruments, 1989
- Far-infrared spectroscopy by synchrotron radiation at the UVSOR FacilityInternational Journal of Infrared and Millimeter Waves, 1986
- Synchrotron radiation for long wavelength spectroscopyInternational Journal of Infrared and Millimeter Waves, 1986
- The electron storage ring as a source of infrared radiationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1985
- Infrared spectroscopy with synchrotron radiationInfrared Physics, 1978
- Synchrotron radiation in the infraredJournal de Physique, 1976
- Synchrotron Radiation as an Infrared SourceApplied Optics, 1973