Optical analysis of liquid mixing in a minibioreactor
- 29 November 2005
- journal article
- research article
- Published by Wiley in Biotechnology and Bioengineering
- Vol. 93 (5) , 906-911
Abstract
A novel optical sensor was used to study mixing and mean circulation time in a model minibioreactor (12.5 mL stirred vessel, equipped with a paddle impeller). Rotational rates in the range of 10–1,000 rpm corresponding to Reynolds number between 14 and 1,350 were studied. Results suggest that depending on the impeller rotational speed, mixing times up to 214 ± 87 s can be reproducibly achieved. The minibioreactor was operated in the transitional regime, and it was determined that the non‐dimensional form for mixing time, NΘM was linearly dependent on Reynolds number. A linear correlation between mean circulation time and the inverse of rotational speed was also determined. The mean circulation time dependence on rotational speed in the 12.5 mL stirred vessel is similar to those found in large‐scale stirred vessels. These results suggest that mixing and circulation times found in large‐scale reactors can be replicated in minibioreactors.Keywords
This publication has 21 references indexed in Scilit:
- Modelling surface aeration rates in shaken microtitre plates using dimensionless groupsChemical Engineering Science, 2005
- Experimental visualization of mixing pathologies in laminar stirred tank bioreactorsChemical Engineering Science, 2005
- Mixing in Unbaffled High-Throughput Experimentation ReactorsIndustrial & Engineering Chemistry Research, 2004
- pH control in microwell fermentations of S. erythraea CA340: influence on biomass growth kinetics and erythromycin biosynthesisBiochemical Engineering Journal, 2003
- Design of a prototype miniature bioreactor for high throughput automated bioprocessingChemical Engineering Science, 2003
- Active volume of mean circulation for stirred tanks agitated with axial impellersChemical Engineering Science, 2000
- Low-cost microbioreactor for high-throughput bioprocessingBiotechnology and Bioengineering, 2000
- Measuring the liquid circulation time in a large gas-liquid contractor by means of a radio pill. Part 2. Circulation time distributionIndustrial & Engineering Chemistry Research, 1987
- New scale-up and design method for stirrer agitated batch mixing vesselsChemical Engineering Science, 1976
- Fluid flow in turbine-stirred, baffled tanks—IChemical Engineering Science, 1964