6 and 15 MeV photon spectra reconstruction using an unfolding depth dose gradient methodology
Authors:
- 1 August 2011
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 2011 (1094687X) , 129-132
Abstract
An accurate knowledge of the spectral distribution emission is essential for precise dose calculations in radiotherapy treatment planning. Reconstruction of photon spectra emitted by medical accelerators from measured depth dose distributions in a water cube is an important tool for commissioning a Monte Carlo treatment planning system. However, the reconstruction problem is an inverse radiation transport function which is poorly conditioned and its solution may become unstable due to small perturbations in the input data. In this paper we present a more stable spectral reconstruction method which can be used to provide an independent confirmation of source models for a given machine without any prior knowledge of the spectral distribution. This technique involves measuring the depth dose curve in a water phantom and applying an unfolding method using Monte Carlo simulated depth dose gradient curves for consecutives mono-energetic beams. We illustrate this theory to calculate a 6 and a 15 MeV photon beam emitted from an Elekta Precise radiotherapy unit using the gradient of depth dose curves in a cube-shaped water tank.Keywords
This publication has 1 reference indexed in Scilit:
- REGULARIZATION TOOLS: A Matlab package for analysis and solution of discrete ill-posed problemsNumerical Algorithms, 1994