Peak identification in capillary isoelectric focusing using the concept of relative peak position as determined by two isoelectric point markers
Authors:
- 15 September 2006
- journal article
- Published by Wiley in Electrophoresis
- Vol. 27 (18) , 3584-3590
Abstract
In CIEF analysis, sample peaks can be identified by their relative peak positions (RPP) that are determined using only two internal pI markers. The two internal pI marker peaks should bracket, as close as possible, the sample peaks. The RPP values of the sample peaks are then calculated using the pI values, peak positions of the two pI markers, and peak position of the sample. Use of this method can effectively compensate for pH gradient distortions that often occur as a result of salts. Also, as shown by the results of this paper, regardless of the linearity of the pH gradient established by the given carrier ampholytes, sample peaks can be identified within an SD of 0.1 pH unit in RPP (<2% RSD) as long as the sample is run using the same carrier ampholytes and maintaining salt concentrations in the range of 0–15 mM.Keywords
This publication has 15 references indexed in Scilit:
- Determination of the isoelectric point of proteins by capillary isoelectric focusingJournal of Chromatography A, 2004
- Preparative-scale, recirculating, pH-biased binary isoelectric trapping separationsElectrophoresis, 2004
- Isoelectric Point Determination of Norovirus Virus-like Particles by Capillary Isoelectric Focusing with Whole Column Imaging DetectionAnalytical Chemistry, 2003
- One-step capillary isoelectric focusing of the proteins in cerebrospinal fluid and serum of patients with neurological disordersJournal of Chromatography A, 2002
- Element of a validation method for MU-B3 monoclonal antibody using an imaging capillary isoelectric focusing systemElectrophoresis, 2002
- Synthetic oligopeptides as isoelectric point markers for capillary isoelectric focusing with ultraviolet absorption detectionElectrophoresis, 2000
- Characterization of Humanized Anti-Tac Monoclonal Antibody by Traditional Separation Techniques and Capillary ElectrophoresisJournal of Liquid Chromatography, 1992
- Separation of the Isoforms of a Monoclonal Antibody by Gel Isoelectric Focusing, High Performance Liquid Chromatography and Capillary Isoelectric FocusingJournal of Liquid Chromatography, 1992
- Adaptation of the equipment for high-performance electrophoresis to isoelectric focusingJournal of Chromatography A, 1985
- Isoelectric points and molecular weights of proteinsJournal of Chromatography A, 1981