High-Speed, High-Resolution Monolithic Capillary LC−MALDI MS Using an Off-Line Continuous Deposition Interface for Proteomic Analysis
Authors:
- 10 March 2005
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 77 (8) , 2323-2331
Abstract
High-speed, high-resolution LC separations, using a poly(styrene-divinylbenzene) monolithic column, have been coupled to MALDI MS and MS/MS through an off-line continuous deposition interface. The LC eluent was mixed with alpha-cyano-4-hydroxycinnamic acid matrix solution and deposited on a MALDI plate that had been precoated with nitrocellulose. Deposition at subatmospheric pressure (80 Torr) formed a 250-microm-wide serpentine trace with uniform width and microcrystalline morphology. The deposited trace was then analyzed in the MS mode using a MALDI-TOF/TOF MS instrument. Continuous deposition allowed interrogation of the separation with a high data sampling rate in the chromatographic dimensions, thus preserving the high resolution of narrow peaks (3-5-s peak width at half-height) of the fast monolithic LC. No extracolumn band broadening due to the deposition process was observed. Over 2000 components were resolved in a 10-min linear gradient separation of the model sample, and 386 unique peptides were identified in the subsequent MS/MS analysis. The continuous deposition interface allows the coupling of high-resolution separations to MALDI MS without degradation in separation efficiency, thus enabling high-throughput proteome analysis.Keywords
This publication has 20 references indexed in Scilit:
- Microdeposition Device Interfacing Capillary Electrochromatography and Microcolumn Liquid Chromatography with Matrix-Assisted Laser Desorption/Ionization Mass SpectrometryAnalytical Chemistry, 2004
- Depth of Proteome IssuesMolecular & Cellular Proteomics, 2004
- Role of Electrons in Laser Desorption/Ionization Mass SpectrometryAnalytical Chemistry, 2003
- A Universal Denoising and Peak Picking Algorithm for LC−MS Based on Matched Filtration in the Chromatographic Time DomainAnalytical Chemistry, 2003
- Exploiting the complementary nature of LC/MALDI/MS/MS and LC/ESI/MS/MS for increased proteome coverageJournal of the American Society for Mass Spectrometry, 2003
- Ion Formation in MALDI: The Cluster Ionization MechanismChemical Reviews, 2003
- Off-Line Coupling of High-Resolution Capillary Electrophoresis to MALDI-TOF and TOF/TOF MSJournal of Proteome Research, 2002
- α-Cyano-4-hydroxycinnamic Acid Affinity Sample Preparation. A Protocol for MALDI-MS Peptide Analysis in ProteomicsAnalytical Chemistry, 2000
- Performance of a Monolithic Silica Column in a Capillary under Pressure-Driven and Electrodriven ConditionsAnalytical Chemistry, 2000
- Direct Analysis and Identification of Proteins in Mixtures by LC/MS/MS and Database Searching at the Low-Femtomole LevelAnalytical Chemistry, 1997