The influence of dilution on the offline measurement of exhaled nitric oxide
- 1 February 2018
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 39 (2) , 025004
Abstract
Objective : Measurement of fractional exhaled nitric oxide (FeNO) is used to determine the presence and severity of eosinophilic airway inflammation in asthma and other wheezing illnesses. The gold standard of online measurement during a single prolonged exhalation is not suitable for use in young children. The international guidelines for offline measurements recommend collection of exhaled gas in an appropriate reservoir for later analysis in young children. The apparatus required for gas collection, however, creates dead space within the system, which may result in sample dilution and hence inaccuracy. Our objective was to investigate the effect such dilution might have on the accuracy of offline FeNO by comparing the results to online results. Approach : Thirty-five adult subjects without respiratory disease underwent online measurement of FeNO and, thereafter, undertook offline FeNO measurements via exhalation into a collection reservoir using one, five or ten inhalation–exhalation cycles. Fifteen of the subjects also exhaled using the five-breath technique via apparatus with additional dead space. An equation incorporating dead space volume and the number of breaths was used to predict the degree of dilution; the predicted results were compared to the measured results. Main results : The median (IQR) FeNO from a one-breath technique (22 (15–28) ppb was not significantly different to online values (19 (12–27) ppb, p = 1.00), but the results from the five-breath technique (11 (4–19) ppb, p < 0.0001), the ten-breath technique (6 (4–15) ppb, p < 0.0001) and the additional dead space experiment (6 (3–8) ppb, p = 0.0006) were significantly lower than online FeNO. Measured values were consistently significantly different to those predicted by the dilution equation, even when incorporating the exact exhaled volume of gas. Significance : Offline FeNO results may be inaccurate when subjects are unable to fill the collection reservoir with a single exhalation, thus the technique may not be suitable for preschool children.Keywords
This publication has 8 references indexed in Scilit:
- A European Respiratory Society technical standard: exhaled biomarkers in lung diseaseEuropean Respiratory Journal, 2017
- Fractional exhaled nitric oxide for the management of asthma in adults: a systematic reviewEuropean Respiratory Journal, 2016
- FeNO and Asthma Treatment in ChildrenMedicine, 2015
- Community-acquired respiratory viruses after lung transplantation: common, sometimes silent, potentially lethalThorax, 2013
- Nasal and Exhaled Nitric Oxide in Allergic RhinitisClinical and Experimental Otorhinolaryngology, 2012
- Nitric oxide synthases: regulation and functionEuropean Heart Journal, 2011
- ATS/ERS Recommendations for Standardized Procedures for the Online and Offline Measurement of Exhaled Lower Respiratory Nitric Oxide and Nasal Nitric Oxide, 2005American Journal of Respiratory and Critical Care Medicine, 2005
- Nasal contribution to exhaled nitric oxide during exhalation against resistance or during breath holdingThorax, 1997