Design of an optimal Ross filter system for x-ray spectra measurements in the range of 8.98–88 keV
Author:
- 1 January 1995
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (1) , 773-775
Abstract
This article describes wide-band Ross filter spectrometers for short-pulse plasma devices. The spectrometer consists of a number of channels, each representing a Ross filter pair in conjunction with x-ray detectors. Sensitivity bands of these channels cover the energy interval of interest without gaps. We consider as optimal the spectrometer whose channels have constant response within their sensitivity ranges and minimal response outside them. The main innovation is the addition of light material into one filter of the Ross pair, which reduces significantly (up to 5 times) the response outside the sensitivity range. Examples are presented of the optimal six-channel (8.98–18.99–29.2–38.92–50.24–67.42–88 keV) spectrometers with three types of detectors: Pb calorimeters, CsI scintillators, and Al2O3 thermoluminescent detectors and filters of readily available materials. The effect of uncertainty in x-ray cross sections and material thicknesses and the effect of filter-scattered photons hitting the detector on the accuracy of measurements are calculated.Keywords
This publication has 3 references indexed in Scilit:
- Study of the x-ray production mechanism of a dense plasma focusJournal of Applied Physics, 1974
- Photon cross sections from 0.1 keV to 1 MeV for elements Z = 1 to Z = 94Atomic Data and Nuclear Data Tables, 1973
- Reevaluation of X-Ray Atomic Energy LevelsReviews of Modern Physics, 1967