Abstract
This study describes a new method for evaluating polyethylene wear in total knee arthroplasty. Since the amount of wear is dependent on a number of variables such as the weight and activity of the patient, it should be estimated based on in vivo measurements. We used a computer vision technique called three-dimensional/two-dimensional (3-D/2-D) matching to perform in vivo assessment using a single-plane radiograph. Using the 3-D/2D matching algorithm we estimated the 3-D position and orientation of each knee implant and then measured the femorotibial distance, which is defined as the shortest perpendicular distance from the tibial tray to the femoral component. The accuracy of the proposed 3-D/2-D matching method was determined by in vitro investigations. The worst errors in in-plane/out-of-plane translations and rotations were 0.20/1.95 mm and 0.17/0.29/spl deg/, respectively. The root-mean-square error in femorotibial distance measurements using real polyethylene inserts was 0.04 mm. Results of in vivo femorotibial distance measurements are also described.Keywords
This publication has 13 references indexed in Scilit:
- Tractive forces during rolling motion of the knee: Implications for wear in total knee replacementJournal of Biomechanics, 1997
- Radiographic wear assessment in a total knee prosthesis: 5- to 9-year follow-up study of 158 kneesThe Journal of Arthroplasty, 1996
- Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopyIEEE Transactions on Biomedical Engineering, 1996
- Recovering the position and orientation of free-form objects from image contours using 3D distance mapsIEEE Transactions on Pattern Analysis and Machine Intelligence, 1995
- A tribological study of a series of retrieved Accord knee explantsMedical Engineering & Physics, 1994
- Failure of the porous-coated anatomic prosthesis in total knee arthroplasty due to severe polyethylene wear.Journal of Bone and Joint Surgery, 1993
- Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prosthesesThe Journal of Bone and Joint Surgery. British volume, 1992
- On recognizing and positioning curved 3-D objects from image contoursIEEE Transactions on Pattern Analysis and Machine Intelligence, 1990
- The accuracy of DLT extrapolation in three-dimensional film analysisJournal of Biomechanics, 1986
- A Simplex Method for Function MinimizationThe Computer Journal, 1965