Abstract
In the current era of total hip arthroplasty (THA), orthopedic surgeons have several fixation options at their disposal. The modern monoblock prosthesis, introduced by Dr. Sir John Charnley, has seen many modifications since its inception in the 1970s and continues to be the most commonly used prosthesis style for primary and revision THA. Proximal modular sleeve technology was introduced in 1967 by Konstantin Sivash, modifying his original 1956 Sivash Stem design. The design is now known as the S-ROM, and although design modifications continue to date, the fundamental structure of the S-ROM remains essentially unchanged. Several other proximal modular prostheses are now currently available for use in THA. Although this similarity in design enables considerable surgical flexibility, it also links their potential for catastrophic failure. This aim of this article was to present a brief history of proximal modularity in THA and to add to the small body of literature regarding catastrophic failure in modular hip implants, including its proposed etiologies including micromotion, fretting and corrosion.Keywords
This publication has 15 references indexed in Scilit:
- Modular Femoral Sleeve and Stem Implant Provides Long-term Total Hip SurvivorshipClinical Orthopaedics and Related Research, 2011
- Corrosion at the Stem-Sleeve Interface of a Modular Titanium Alloy Femoral Component as a Reason for Impaired DisengagementThe Journal of Arthroplasty, 2011
- Study of Factors Affecting Taper Joint Failures in Modular Hip Implant Using Finite Element Modelling2010
- Metal release and corrosion effects of modular neck total hip arthroplastyInternational Orthopaedics, 2009
- In vivo severe corrosion and hydrogen embrittlement of retrieved modular body titanium alloy hip‐implantsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- Fretting crevice corrosion of stainless steel stem–CoCr femoral head connections: Comparisons of materials, initial moisture, and offset lengthJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- Differences in the fretting corrosion of metal–metal and ceramic–metal modular junctions of total hip replacementsJournal of Orthopaedic Research, 2004
- Head—neck modularity for total hip arthroplastyThe Journal of Arthroplasty, 1996
- Design-related fretting wear in modular neck hip prosthesisJournal of Biomedical Materials Research, 1996
- Laboratory testing of displacement and load induced frettingTribology International, 1995