Single pencil beam benchmark of a module for Monte Carlo simulation of proton transport in the PENELOPE code
Authors:
Open Access
- 20 November 2020
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (1) , 456-476
Abstract
Background and purpose: PENH is a recently coded module for simulation of proton transport in conjunction with the Monte Carlo code PENELOPE. PENELOPE applies class II simulation to all type of interactions, in particular, to elastic collisions. PENH uses calculated differential cross sections for proton elastic collisions that include electron screening effects as well as nuclear structure effects. Proton‐induced nuclear reactions are simulated from information in the ENDF‐6 database or from alternative nuclear databases in ENDF format. The purpose of this work is to benchmark this module by simulating absorbed dose distributions from a single finite spot size proton pencil beam in water.Materials and methods: Monte Carlo simulations with PENH are compared with simulation results from TOPAS Monte Carlo (v3.1p2) and RayStation Monte Carlo (v6). Different beam models are examined in terms of mean energy and energy spread to match the measured profiles. The phase‐space file is derived from experimental measurements. Simulated absorbed dose distributions are compared to experimental data obtained with the ionization chamber array MatriXX 2D detector (IBA Dosimetry) in a water tank. The experiments were conducted with a clinical IBA pencil beam scanning dedicated nozzle. In all simulations a Fermi‐Eyges phase‐space representation of a single finite spot size proton pencil beam is used.Results: In general, there is a good agreement between simulated results and experimental data up to a distance of 3 cm from the central axis. In the core region (region where the dose is more than 10% of the maximum dose) PENH shows, overall, the smallest deviations from experimental data, with the largest radial rms (root mean square) smaller than 0.2. The results achieved by TOPAS and RayStation in that region are very close to those of PENH. For the halo region, that is the area of the dose distribution outside the core region reaching down to 0.01% of the maximum intensity, the largest rms achieved by TOPAS is always smaller than 0.5, yielding better results than the rest of the codes.Conclusion: The physics modeling of the PENELOPE/PENH code yields results consistent with measurements in the dose range relevant for proton therapy. The discrepancies between PENH appearing at distances larger than 3 cm from the central‐beam axis are accountable to the lack of neutron simulation in this code. In contradistinction, TOPAS has a better agreement with experimental data at large distances from the central‐beam axis because of the simulation of neutrons.Keywords
This publication has 34 references indexed in Scilit:
- A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profilesPhysics in Medicine & Biology, 2013
- TOPAS: An innovative proton Monte Carlo platform for research and clinical applicationsMedical Physics, 2012
- Range uncertainties in proton therapy and the role of Monte Carlo simulationsPhysics in Medicine & Biology, 2012
- Golden beam data for proton pencil-beam scanningPhysics in Medicine & Biology, 2012
- Configuration of the electron transport algorithm of PENELOPE to simulate ion chambersPhysics in Medicine & Biology, 2006
- An analytical approximation of the Bragg curve for therapeutic proton beamsMedical Physics, 1997
- An algorithm for Monte Carlo simulation of coupled electron-photon transportNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1997
- Numerical Computation of Multivariate Normal ProbabilitiesJournal of Computational and Graphical Statistics, 1992
- Multiple Scattering with Energy LossPhysical Review B, 1948
- Cosmic-Ray TheoryReviews of Modern Physics, 1941