Critical appraisal of the CLSI guideline EP09c “measurement procedure comparison and bias estimation using patient samples”
- 19 August 2024
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 63 (3) , 507-514
Abstract
Background: In laboratory setting evaluating the agreement between two measurement methods is a very frequent practice. Unfortunately, the guidelines to refer to are not free from criticisms from a statistical methodological point of view. We reviewed the Clinical and Laboratory Standards Institute guideline EP09c, 3rd ed. pointing out some drawbacks and some aspects that have not been well defined, leaving situations of uncertainty and/or of excessive subjectivity in the judgement. Content: We have stressed the need of having replicates to estimate the systematic and the proportional biases of the measurement methods to be compared. Indeed, unequal variance of the two measurement methods gives a slope and intercept of the regression between the difference and the mean of the two values of the measurement methods to be compared that can be absolutely calculated from their means, their variances and their correlation coefficient. So, it is not possible to disentangle true from spurious biases. For laboratory professionals we have developed a worked exemplification of an agreement assessment. Summary: We have stressed the need of other approaches than the classic Bland and Altman method to calculate the systematic and proportional biases of two measurement methods compared for their agreement in a study with replicates.Keywords
This publication has 30 references indexed in Scilit:
- The Bland-Altman range of agreement: Exact interval procedure and sample size determinationComputers in Biology and Medicine, 2018
- The role of external quality assessment in the verification of in vitro medical diagnostics in the traceability eraClinical Biochemistry, 2018
- Comparing two measurement systems using the probability of agreement web appQuality Engineering, 2017
- Comparing Clinical Measurement Methods2010
- Why Bland–Altman plots should use X, not (Y+X)/2 when X is a reference methodStatistics in Medicine, 2007
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- Reference measurement systems in clinical chemistryClinica Chimica Acta, 2002
- Comparing methods of measurement: why plotting difference against standard method is misleadingThe Lancet, 1995
- Estimation of the linear relationship between the measurements of two methods with proportional errorsStatistics in Medicine, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986