Acoustic noise‐optimized verse pulses
Authors:
Open Access
- 26 October 2010
- journal article
- imaging methodology---note
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 64 (5) , 1446-1452
Abstract
Variable‐rate selective excitation RF pulses modulate the slice selection gradients during RF transmission, especially to reduce the total RF power. Amplitude‐modulated slice selection gradients can lead to increased gradient noise, in particular in high‐field MRI where variable‐rate selective excitation techniques are often used. In this work, an algorithm is presented that calculates a variable‐rate selective excitation pulse modulation from given RF pulses with constant slice selection gradient. The algorithm avoids the known acoustic resonance frequencies of the gradient system to minimize sound pressure levels. It was tested with four different slice‐selective RF pulse shapes (Sinc, Gaussian, and two Shinnar‐LeRoux). Sound measurements revealed a reduction of the mean sound pressure level by up to 13dB, and simultaneously, the specific absorption rate was reduced by 55%. Magn Reson Med, 2010.Keywords
This publication has 14 references indexed in Scilit:
- High-flip-angle slice-selective parallel RF transmission with 8 channels at 7TJournal of Magnetic Resonance, 2008
- Design and optimization for variable rate selective excitation using an analytic RF scaling functionJournal of Magnetic Resonance, 2007
- Variable‐rate selective excitation for rapid MRI sequencesMagnetic Resonance in Medicine, 2004
- New FOCI pulses with reduced radiofrequency power requirementsJournal of Magnetic Resonance Imaging, 2004
- Making MRI QuieterMagnetic Resonance Imaging, 2002
- Characterization and prediction of gradient acoustic noise in MR imagersMagnetic Resonance in Medicine, 1997
- Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)IEEE Transactions on Medical Imaging, 1991
- A reduced power selective adiabatic spin‐echo pulse sequenceMagnetic Resonance in Medicine, 1991
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988
- Power deposition in whole-body NMR imagingMedical Physics, 1981