Abstract
A disposable dual screen‐printed electrode was designed and evaluated for the direct and simultaneous determination of chromium(VI) and chromium(III) in batch and flow analysis. For this purpose a carbon screen‐printed bielectrode was modified with a graphite‐epoxy composite which contains diphenylcarbazide and another graphite‐epoxy composite prepared with di(2‐hydroxyphenylimino)ethane. The principal analytical parameters of the potentiometric responses in batch and flow analysis were optimized and calculated. The bisensor shows optimal detection limits of 7.7×10−7 M and 7.9×10−6 M in static mode, and 9.0×10−7 M and 1.2×10−5 M in flow analysis for hexavalent and trivalent chromium, respectively. This potentiometric device was employed with successful results to determine both chromium species in leachates from municipal solid waste landfills.

This publication has 23 references indexed in Scilit: