Abstract
To determine whether the NiTiNol frame of a novel hernia repair device utilizing polypropylene mesh for inguinal hernioplasty remains stable and intransient without fixation after a minimum of 6 months. Twenty patients had 27 inguinal hernias repaired using a novel hernia repair device that has a NiTiNol frame without any fixation. Initial single-view, postoperative X-rays were compared with a second X-ray obtained at least 6 months later. The NiTiNol frame, which can be easily visualized on a plain X-ray, was measured in 2 dimensions, as were anatomic landmarks. The measurements obtained and the appearances of the 2 X-rays were compared to determine the percentage of change in device size and device stability with regard to device location and shape. There were minimal changes noted between the 2 sets of measurements obtained with an overall trend towards a slight increase in the size of the hernia repair device. The devices demonstrated intransience of position and stability of shape. The NiTiNol frame of a novel hernia repair device utilizing polypropylene mesh exhibits radiographic evidence of size and shape stability and intransience of position without fixation when used in inguinal hernioplasty after a minimum follow-up of 6 months.