Validation of the doubly labeled water method using off-axis integrated cavity output spectroscopy and isotope ratio mass spectrometry
- 1 February 2018
- journal article
- research article
- Published by American Physiological Society in American Journal of Physiology-Endocrinology and Metabolism
- Vol. 314 (2) , E124-E130
Abstract
When the doubly labeled water (DLW) method is used to measure total daily energy expenditure (TDEE), isotope measurements are typically performed using isotope ratio mass spectrometry (IRMS). New technologies, such as off-axis integrated cavity output spectroscopy (OA-ICOS) provide comparable isotopic measurements of standard waters and human urine samples, but the accuracy of carbon dioxide production (V̇co2) determined with OA-ICOS has not been demonstrated. We compared simultaneous measurement V̇co2 obtained using whole-room indirect calorimetry (IC) with DLW-based measurements from IRMS and OA-ICOS. Seventeen subjects (10 female; 22 to 63 yr) were studied for 7 consecutive days in the IC. Subjects consumed a dose of 0.25 g H218O (98% APE) and 0.14 g 2H2O (99.8% APE) per kilogram of total body water, and urine samples were obtained on days 1 and 8 to measure average daily V̇co2 using OA-ICOS and IRMS. V̇co2 was calculated using both the plateau and intercept methods. There were no differences in V̇co2 measured by OA-ICOS or IRMS compared with IC when the plateau method was used. When the intercept method was used, V̇co2 using OA-ICOS did not differ from IC, but V̇co2 measured using IRMS was significantly lower than IC. Accuracy (~1–5%), precision (~8%), intraclass correlation coefficients ( R = 0.87–90), and root mean squared error (30–40 liters/day) of V̇co2 measured by OA-ICOS and IRMS were similar. Both OA-ICOS and IRMS produced measurements of V̇co2 with comparable accuracy and precision compared with IC.Keywords
Funding Information
- HHS | National Institutes of Health (NIH) (R44 DK093362, P30 DK048520)
- Foundation for the National Institutes of Health (FNIH) (UL1 RR025780)
This publication has 23 references indexed in Scilit:
- Measurement of δ18O, δ17O, and 17O-excess in Water by Off-Axis Integrated Cavity Output Spectroscopy and Isotope Ratio Mass SpectrometryAnalytical Chemistry, 2013
- Validity of the doubly labeled water method for estimating CO2production in mice under different nutritional conditionsAmerican Journal of Physiology-Endocrinology and Metabolism, 2013
- Direct Analysis of δ2H and δ18O in Natural and Enriched Human Urine Using Laser-Based, Off-Axis Integrated Cavity Output SpectroscopyAnalytical Chemistry, 2012
- Spectral contaminant identifier for off-axis integrated cavity output spectroscopy measurements of liquid water isotopesReview of Scientific Instruments, 2012
- Doubly labeled water analysis using cavity ring‐down spectroscopyRapid Communications in Mass Spectrometry, 2010
- A new approach for flow-through respirometry measurements in humansAmerican Journal of Physiology-Regulatory, Integrative and Comparative Physiology, 2010
- Comparison of Short Term Indirect Calorimetry and Doubly Labeled Water Method for the Assessment of Energy Expenditure in Preterm InfantsNeonatology, 1991
- Assessment of Energy Expenditure and Fuel Utilization in ManAnnual Review of Nutrition, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Theory of use of the turnover rates of body water for measuring energy and material balanceJournal of Theoretical Biology, 1966