Miniature high-resolution thermometer for low-temperature applications
Authors:
- 1 September 2001
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 72 (9) , 3600-3604
Abstract
We report on a new miniature, high-resolution, susceptibilitythermometer that employs La x Gd 1−x Cl 3 as the paramagnetic material for temperature measurement below 4 K. The thermometer utilizes small permanent disk magnets to provide the required magnetic field, and has a total mass of only 2 g. The thermometer shows a resolution of 0.2 nK in a 1 Hz bandwidth and an absolute drift rate less than 50 fK/s at 2.2 K. In this article, we describe the thermometer design, assembly procedure, and experimental test results, including its thermal time constant and susceptibility to external magnetic fields.Keywords
This publication has 14 references indexed in Scilit:
- A new small nano-Kelvin resolution thermometer for low-temperature experimentsIEEE Transactions on Instrumentation and Measurement, 2000
- Superconducting magnetic shields for SQUID applicationsReview of Scientific Instruments, 1999
- The fluctuation-imposed limit for temperature measurementJournal of Low Temperature Physics, 1997
- A high-resolution susceptibility thermometer for use near 2 KCzechoslovak Journal of Physics, 1996
- High resolution thermometry for the confined helium experimentCzechoslovak Journal of Physics, 1996
- Heat Capacity and Thermal Relaxation of Bulk Helium very near the Lambda PointPhysical Review Letters, 1996
- Temperature fluctuations in the canonical ensemblePhysical Review Letters, 1992
- A very high resolution thermometer for use below 7 KPhysica B+C, 1981
- The magnetic barkhausen effectC R C Critical Reviews in Solid State Sciences, 1976
- Irreversibility and Generalized NoisePhysical Review B, 1951