X-ray powder diffraction characterization of iron microparticles on a Bruker SMART1000 single-crystal X-ray diffractometer
- 1 December 2009
- journal article
- Published by Cambridge University Press (CUP) in Powder Diffraction
- Vol. 24 (4) , 347-350
Abstract
A method to characterize iron oxide microparticles by microquantity X-ray powder diffraction is presented. The method employs small sections of acrylic tubing and double sided Mylar ® tape that fits over a standard 170 mm collimator for a Bruker-AXS SMART 1000 or APEXII three-circle single-crystal diffractometer (Mo X-ray tube). The tubing will hold and position a sample that is placed on the double-sided Mylar ® tape and allow for rapid specimen mounting/dismounting and data collection. The method is simple, portable, and readily adapted to a variety of single-crystal X-ray diffractometers.Keywords
This publication has 6 references indexed in Scilit:
- SupermagnetismJournal of Physics D: Applied Physics, 2008
- Magnetic Iron Oxide Nanoparticles: Synthesis, Stabilization, Vectorization, Physicochemical Characterizations, and Biological ApplicationsChemical Reviews, 2008
- New method for separation of magnetite from rock samples for oxygen isotope analysisEuropean Journal of Mineralogy, 2007
- A novel approach to micro-sample X-ray powder diffraction using nylon loopsJournal of Applied Crystallography, 2003
- Preparation of aqueous magnetic liquids in alkaline and acidic mediaIEEE Transactions on Magnetics, 1981
- IMPROVED MOUNT FOR POWDERED SPECIMENS USED ON THE GEIGER‐COUNTER X‐RAY SPECTROMETER*Journal of the American Ceramic Society, 1949