A Wireless Integrated Circuit for 100-Channel Charge-Balanced Neural Stimulation
Authors:
- 24 November 2009
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. 3 (6) , 405-414
Abstract
The authors present the design of an integrated circuit for wireless neural stimulation, along with benchtop and in - vivo experimental results. The chip has the ability to drive 100 individual stimulation electrodes with constant-current pulses of varying amplitude, duration, interphasic delay, and repetition rate. The stimulation is performed by using a biphasic (cathodic and anodic) current source, injecting and retracting charge from the nervous system. Wireless communication and power are delivered over a 2.765-MHz inductive link. Only three off-chip components are needed to operate the stimulator: a 10-nF capacitor to aid in power-supply regulation, a small capacitor (< 100 pF) for tuning the coil to resonance, and a coil for power and command reception. The chip was fabricated in a commercially available 0.6- mum 2P3M BiCMOS process. The chip was able to activate motor fibers to produce muscle twitches via a Utah Slanted Electrode Array implanted in cat sciatic nerve, and to activate sensory fibers to recruit evoked potentials in somatosensory cortex.Keywords
This publication has 24 references indexed in Scilit:
- An Integrated System for Simultaneous, Multichannel Neuronal Stimulation and RecordingIEEE Transactions on Circuits and Systems I: Regular Papers, 2007
- A 232-Channel Epiretinal Stimulator ASICIEEE Journal of Solid-State Circuits, 2007
- Designing Efficient Inductive Power Links for Implantable Devices2007
- A Fully Integrated Fail-safe Stimulator Output Stage Dedicated to FES Stimulation2007
- A Low-Power Integrated Circuit for a Wireless 100-Electrode Neural Recording SystemIEEE Journal of Solid-State Circuits, 2006
- A matched biphasic microstimulator for an implantable retinal prosthetic device2004
- CMOS neurostimulation ASIC with 100 channels, scaleable output, and bidirectional radio-frequency telemetryIEEE Transactions on Biomedical Engineering, 2001
- An inherently linear and compact MOST-only current division techniqueIEEE Journal of Solid-State Circuits, 1992
- Imbalanced biphasic electrical stimulation: Muscle tissue damageAnnals of Biomedical Engineering, 1990
- The Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve StimulationIEEE Transactions on Biomedical Engineering, 1983