An Isothermal Peltier Cell Calorimeter For Measuring the Magnetocaloric Effect
- 21 May 2007
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Magnetics
- Vol. 43 (6) , 2764-2766
Abstract
We have developed a method for isothermal calorimetry for the determination of the entropy change in magnetic materials. Heat absorbed and released by the sample at constant temperature is measured by means of Peltier cells, used in an active way to simultaneously measure the heat current and keep constant the sample temperature. In the paper we describe the method, the Peltier cell calibration and the test apparatus. The results obtained on a Gadolinium sample at room temperature under a magnetic field up to 0.9 T are discussed. The comparison with other methods demonstrates the validity of the proposed approach to fulfil the isothermal conditions (DeltaTles5 mK) and to obtain a faster time responseKeywords
This publication has 5 references indexed in Scilit:
- Recent developments in magnetocaloric materialsReports on Progress in Physics, 2005
- Isothermal reaction calorimetry as a tool for kinetic analysisThermochimica Acta, 2004
- A high-sensitivity differential scanning calorimeter with magnetic field for magnetostructural transitionsReview of Scientific Instruments, 2003
- Development of an isothermal titration microcalorimetric system with digital control and dynamic power Peltier compensation. I. Description and basic performanceReview of Scientific Instruments, 2000
- Magnetocaloric effect from indirect measurements: Magnetization and heat capacityJournal of Applied Physics, 1999