Diamond Beam-Position Monitor for Undulator Radiation and Tests at the Tristan Super Light Facility

Authors:
  • HSH. Sakae
    H. Sakae
  • HAH. Aoyagi
    H. Aoyagi
  • HKH. Kimura
    H. Kimura
  • TOT. Ohata
    T. Ohata
  • HSH. Shiwaku
    H. Shiwaku
  • SYS. Yamamoto
    S. Yamamoto
  • HSH. Sugiyama
    H. Sugiyama
  • KTK. Tanabe
    K. Tanabe
  • KKK. Kobaski
    K. Kobaski
  • HKH. Kitamura
    H. Kitamura

Abstract
A photon beam-position monitor using a diamond foil as a position-sensitive device has been developed for use on insertion-device beamlines of third-generation synchrotron radiation facilities such as SPring-8, and was tested on the undulator beamline of the Tristan Super Light Facility at KEK. The beam test results show that the diamond monitor can be operated in a photoconductive mode like a semiconductor detector. It has a linear working range of about +/-1 mm and a position sensitivity of less than 3 mum. The stability of the monitor was confirmed by continuous operation under low photon beam intensity conditions.