Improvement of the Bone-Pin Interface Strength in Osteoporotic Bone with Use of Hydroxyapatite-Coated Tapered External-Fixation Pins
Authors:
- Antonio MoroniAMAntonio Moroni
- Cesare FaldiniCFCesare Faldini
- Stefano MarchettiSMStefano Marchetti
- Mario MancaMMMario Manca
- Vincenzo ConsoliVCVincenzo Consoli
- Sandro GianniniSGSandro Giannini
- 1 May 2001
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 83 (5) , 717-721
Abstract
Achieving adequate fixation strength in osteoporotic bone is a challenge. In this study, we examined the use of hydroxyapatite-coated tapered external-fixation pins for the fixation of wrist fractures in patients with osteoporosis. Twenty female patients with osteoporosis and a fracture of the wrist were divided into two paired groups and randomized to receive either standard tapered pins or hydroxyapatite-coated tapered pins. Two pins were inserted in the distal part of the radius, two pins were inserted in the second metacarpal, and an external fixation device was mounted. All fixation devices were removed six weeks after surgery. The mean pin-insertion torque (and standard deviation) was 461 +/- 254 Nmm in the group managed with standard pins and 332 +/- 176 Nmm in the group managed with hydroxyapatite-coated pins (p = 0.01). The mean pin-extraction torque was 191 +/- 155 Nmm in the group managed with standard pins and 600 +/- 214 Nmm in the group managed with hydroxyapatite-coated pins (p < 0.0001, power 95%). The mean extraction torque was lower than the corresponding insertion torque at each pin position in the group managed with standard pins (p < 0.05), whereas the mean extraction torque was higher than the corresponding insertion torque at each pin position in the group managed with hydroxyapatite-coated pins (p = 0.001). Two patients managed with standard pins and no patient managed with hydroxyapatite-coated pins had a pin-track infection. Pain during pin removal did not differ between the two groups. The present study showed that hydroxyapatite-coated tapered external-fixation pins provided improved fixation in the treatment of wrist fractures in patients with osteoporosis.Keywords
This publication has 8 references indexed in Scilit:
- Redisplaced unstable fractures of the distal radius: A randomised, prospective study of bridging versus non-bridging external fixationThe Journal of Bone and Joint Surgery. British volume, 1998
- A Comparison of Hydroxyapatite-Coated, Titanium-Coated, and Uncoated Tapered External-Fixation Pins. An in Vivo Study in Sheep*Journal of Bone and Joint Surgery, 1998
- Hydroxyapatite Coating Of Threaded Pins Enhances FixationThe Journal of Bone and Joint Surgery. British volume, 1997
- Biomechanical, Scanning Electron Microscopy, and Microhardness Analyses of the Bone-Pin Interface in Hydroxyapatite Coated Versus Uncoated PinsJournal of Orthopaedic Trauma, 1997
- Improvement of the Bone-Pin Interface with Hydroxyapatite Coating: An In Vivo Long-Term Experimental StudyJournal of Orthopaedic Trauma, 1996
- Hydroxyapatite Coated External Fixation PinsClinical Orthopaedics and Related Research, 1996
- CORTICAL BONE REACTIONS AT THE INTERFACE OF EXTERNAL FIXATION HALF-PINS UNDER DIFFERENT LOADING CONDITIONS1993
- Chronic osteomyelitis in pin tracks.Journal of Bone and Joint Surgery, 1984