Intravascular electric impedance spectroscopy of atherosclerotic lesions using a new impedance catheter system
Authors:
- T. SüselbeckTST. Süselbeck
- H. ThieleckeHTH. Thielecke
- J. KöchlinJKJ. Köchlin
- S. ChoSCS. Cho
- I. WeinschenkIWI. Weinschenk
- J. MetzJMJ. Metz
- M. BorggrefeMBM. Borggrefe
- K. K. HaaseKHK. K. Haase
- 1 April 2005
- journal article
- research article
- Published by Springer Nature in Basic Research in Cardiology
- Vol. 100 (5) , 446-452
Abstract
Newer techniques are required to identify atherosclerotic lesions that are prone to rupture. Electric impedance spectroscopy (EIS) can characterize biological tissues by measuring the electrical impedance over a frequency range. We tested a newly designed intravascular impedance catheter (IC) by measuring the impedance of different stages of atherosclerosis induced in an animal rabbit model. Six female New Zealand White rabbits were fed for 17 weeks with a 5% cholesterol–enriched diet to induce early forms of atherosclerotic plaques. All aortas were prepared from the aortic arch to the renal arteries and segments of 5–10 mm were marked by ink spots. A balloon catheter system with an integrated polyimide–based microelectrode structure was introduced into the aorta and the impedance was measured at each spot by using an impedance analyzer. The impedance was measured at frequencies of 1 kHz and 10 kHz and compared with the corresponding histomorphometric data of each aortic segment. Forty–four aortic segments without plaques and 48 segments with evolving atherosclerotic lesions could be exactly matched by the histomorphometric analysis. In normal aortic segments (P0) the change of the magnitude of impedance at 1 kHz and at 10 kHz (|Z|1 kHz – |Z|10 kHz, = ICF) was 208.5 ± 357.6 Ω. In the area of aortic segments with a plaque smaller than that of the aortic wall diameter (PI), the ICF was 137.7 ± 192.8 Ω. (P 0 vs. P I; p = 0.52), whereas in aortic segments with plaque formations larger than the aortic wall (PII) the ICF was significantly lower –22.2 ± 259.9 Ω. (P0 vs. PII; p = 0.002). Intravascular EIS could be successfully performed by using a newly designed microelectrode integrated onto a conventional coronary balloon catheter. In this experimental animal model atherosclerotic aortic lesions showed significantly higher ICF in comparison to the normal aortic tissue.Keywords
This publication has 19 references indexed in Scilit:
- Incidence of High-Strain Patterns in Human Coronary ArteriesCirculation, 2004
- Coronary Plaque Classification With Intravascular Ultrasound Radiofrequency Data AnalysisCirculation, 2002
- Identification of Atherosclerotic Plaque Components With Intravascular Ultrasound Elastography In VivoCirculation, 2002
- Thermal Heterogeneity Within Human Atherosclerotic Coronary Arteries Detected In VivoCirculation, 1999
- Development of an intravascular impedance catheter for detection of fatty lesions in arteriesIEEE Transactions on Medical Imaging, 1997
- The dielectric properties of biological tissues: I. Literature surveyPhysics in Medicine & Biology, 1996
- Coronary Plaque DisruptionCirculation, 1995
- Electrical impedance of layered atherosclerotic plaques on human aortasIEEE Transactions on Biomedical Engineering, 1992
- The Pathogenesis of Coronary Artery Disease and the Acute Coronary SyndromesNew England Journal of Medicine, 1992
- Angiographic progression of coronary artery disease and the development of myocardial infarctionJACC, 1988