Electrosurgery: heating, sparking and electrical arcs
Authors:
Open Access
- 1 September 2024
- journal article
- research article
- Published by Galenos Yayinevi in Human Reproduction Open
- Vol. 16 (3) , 281-290
Abstract
The translation of impedance (R), current (I), and voltage (V) into tissue effects and the understanding of the settings of electrosurgical units is not obvious if judged by the many questions during live surgery. Below 200 V, the current heats the tissue until the steam of boiling stops the current. Thus, slower heating, because of less energy or a larger contact area, results in deeper coagulation. Above 200 V and a duty cycle (per cent of time electricity is delivered) of >50% (yellow pedal), sparks become electric arcs, and the heat causes the explosion of superficial cells, i.e. cutting. With higher voltages, cutting is associated with coagulation, i.e. blended current. With even higher voltages and a duty cycle <10% preventing arching, only coagulation occurs (blue pedal; forced coagulation). Voltage being crucially important for tissue effects, newer electrosurgical units deliver a constant voltage and limit the energy output (Maximal Watts: W=I*V= joules/sec). Unfortunately, the electrosurgical units indicate the combination of voltage and duty cycles as a force of cutting (pure cutting or blended) or coagulation (soft, forced or spray) current. It is important that the surgeon understands whether electrosurgical units control voltages or output, as well as the electrical basics of the different settings and programs used.Keywords
This publication has 33 references indexed in Scilit:
- Efficacy of electrocoagulation hemostasis: a study on the optimal usage of the very-low-voltage modeSurgical Endoscopy, 2022
- Robotic stray energy with constant-voltage versus constant-power regulating electrosurgical generatorsSurgical Endoscopy, 2022
- Temperature profile and residual heat of monopolar laparoscopic and endoscopic dissection instrumentsSurgical Endoscopy, 2021
- Acoustic signal analysis of instrument–tissue interaction for minimally invasive interventionsInternational Journal of Computer Assisted Radiology and Surgery, 2020
- Single-incision laparoscopic surgery increases the risk of unintentional thermal injury from the monopolar “Bovie” instrument in comparison with traditional laparoscopySurgical Endoscopy, 2016
- Insulation failure in electrosurgery instrumentation: a prospective evaluationSurgical Endoscopy, 2016
- Unintended stray energy from monopolar instruments: beware the dispersive electrode cordSurgical Endoscopy, 2015
- Fundamentals of Electrosurgery Part I: Principles of Radiofrequency Energy for Surgery2012
- Insulation failure in laparoscopic instrumentsSurgical Endoscopy, 2009
- Principles of Electrosurgery, Laser, and Argon Plasma Coagulation with Particular Regard to Endoscopy2008