Digitized Infrared Segmental Thermometry: Time Requirements for Stable Recordings
Authors:
- Jean P. BoucherJBJean P. Boucher
- Alain S. ComtoisACAlain S. Comtois
- 1 July 2006
- journal article
- Published by National University of Health Sciences in Journal of Manipulative and Physiological Therapeutics
- Vol. 29 (6) , 468.e1-468.e10
Abstract
Digitized infrared segmental thermometry (DIST) is a method for measuring and recording skin surface body temperatures. The project evaluated the required length of time for patients to acclimatize their core body temperature to ambient conditions to obtain stable DIST readings. Seventeen subjects were allowed a 20-minute acclimatizing period in a temperature-controlled room. The bilateral DIST temperature was measured with thermistors in combination with infrared cameras (IRCs) at the C4 and L4 levels. All IRC temperatures were recorded after a 20-minute stabilization period. The room temperature and relative humidity were recorded throughout all trials. The acclimatization trend was computed from the 20- to 24-minute period for the IRCs, and the acclimatization trend was computed continuously for a total of 30 minutes (at 2-minute intervals) for 5 days. We discovered a stabilization trend in the early trial stages, with the thermistor recordings between 8 and 16 minutes. The IRC trend was also conclusive for the core temperature requirements. This study determined a core body temperature acclimatization trend tested among patients using thermistor recordings in a controlled environment. Based on these findings, we recommend acclimatization in a temperature- and humidity-controlled environment for a minimum 8-minute period, followed by an 8-minute maximum recording period with the patient in a prone position to obtain accurate DIST recordings.Keywords
This publication has 7 references indexed in Scilit:
- Validity of Infrared Thermal Measurements of Segmental Paraspinal Skin Surface TemperatureJournal of Manipulative and Physiological Therapeutics, 2006
- Stability of paraspinal thermal patterns during acclimationJournal of Manipulative and Physiological Therapeutics, 2004
- Reliability and normal values for measuring the skin temperature of the hand with an infrared tympanic thermometer: A pilot studyJournal of Hand Therapy, 1999
- Thermatomal changes in cervical disc herniationsYonsei Medical Journal, 1999
- Quantification of thermal asymmetryJournal of Neurosurgery, 1988
- Quantification of thermal asymmetryJournal of Neurosurgery, 1988
- Normal thermographic standards for the cervical spine and upper extremitiesSkeletal Radiology, 1984