Abstract
A new audiometric device to determine the location of the apical foramen of teeth undergoing endodontic treatment according to “the relative value method” was tested in in vitro experiments. In 96.5% of the 350 human teeth used in the experiment, the location of the apical foramen was determined within acceptable limits (0–0.5 mm). In 60 teeth, where the measurements were based on an intermittent audio signal, a point about 1 mm short of the apical foramen was located. The results were considered promising and a clinical evaluation of the device will be carried out.Keywords
This publication has 9 references indexed in Scilit:
- THE ELECTRONIC DETERMINATION OF WORKING LENGTHDental Clinics of North America, 1992
- Poster clinic #4 — Clinical evaluation in electronic measuring of root canal length using several devicesJournal of Endodontics, 1992
- An experimental study of the principle of electronic root canal measurementJournal of Endodontics, 1987
- A new method and apparatus for measuring root canal length.The Journal of Nihon University School of Dentistry, 1986
- A model for demonstrating an electronic canallength measuring deviceJournal of Endodontics, 1983
- New principle and method for measuring the root canal lengthJournal of Endodontics, 1983
- Reliability of electronic measuring devices in endodontic therapyJournal of Endodontics, 1975
- A study of radiographic longitudinal distortion of anterior teeth using the paralleling techniqueOral Surgery, Oral Medicine, Oral Pathology, 1966
- New Method for Measuring the Length of the Root CanalJournal of Dental Research, 1962