Precision grid and hand motion for accurate needle insertion in brachytherapy
Authors:
- Carl S. McGillCMCarl S. McGill
- Jonathon A. SchwartzJSJonathon A. Schwartz
- Jason Z. MooreJMJason Z. Moore
- Patrick W. McLaughlinPMPatrick W. McLaughlin
- Albert J. ShihASAlbert J. Shih
- 29 July 2011
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 38 (8) , 4749-4759
Abstract
In prostate brachytherapy, a grid is used to guide a needle tip toward a preplanned location within the tissue. During insertion, the needle deflects en route resulting in target misplacement. In this paper, 18-gauge needle insertion experiments into phantom were performed to test effects of three parameters, which include the clearance between the grid hole and needle, the thickness of the grid, and the needle insertion speed. Measurement apparatus that consisted of two datum surfaces and digital depth gauge was developed to quantify needle deflections. The gauge repeatability and reproducibility (GR&R) test was performed on the measurement apparatus, and it proved to be capable of measuring a 2 mm tolerance from the target. Replicated experiments were performed on a 2(3) factorial design (three parameters at two levels) and analysis included averages and standard deviation along with an analysis of variance (ANOVA) to find significant single and two-way interaction factors. Results showed that grid with tight clearance hole and slow needle speed increased precision and accuracy of needle insertion. The tight grid was vital to enhance precision and accuracy of needle insertion for both slow and fast insertion speed; additionally, at slow speed the tight, thick grid improved needle precision and accuracy. In summary, the tight grid is important, regardless of speed. The grid design, which shows the capability to reduce the needle deflection in brachytherapy procedures, can potentially be implemented in the brachytherapy procedure.Keywords
This publication has 11 references indexed in Scilit:
- Observations on rotating needle insertions using a brachytherapy robotPhysics in Medicine & Biology, 2007
- Needle insertion into soft tissue: A surveyMedical Engineering & Physics, 2006
- In vivo motion and force measurement of surgical needle intervention during prostate brachytherapyMedical Physics, 2006
- Force Modeling for Needle Insertion Into Soft TissueIEEE Transactions on Biomedical Engineering, 2004
- Needle insertion modeling and simulationIEEE Transactions on Robotics and Automation, 2003
- Complications Following Permanent Prostate BrachytherapyEuropean Urology, 2002
- Brachytherapy for carcinoma of the prostate: Techniques, patient selection, and clinical outcomesSeminars in Radiation Oncology, 2002
- Dosimetric effects of needle divergence in prostate seed implant using and radioactive seedsMedical Physics, 2000
- Source placement error for permanent implant of the prostateMedical Physics, 1997
- Brachytherapy for prostate cancerCA: A Cancer Journal for Clinicians, 1995